Omron NE-C29-E instruction manual Troubleshooting, Symptom Cause Remedy

Page 22

8. Troubleshooting

Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions.

Symptom

Cause

Remedy

 

 

 

Nothing happens when the

Is the power plug plugged

Check that the plug is

inserted in an electrical

power switch is pressed.

into an electrical outlet?

outlet. Unplug then reinsert

 

 

the plug if necessary.

 

 

 

 

Is there medication in the

Fill the medication tank.

 

medication tank?

 

 

 

 

 

 

Is there too much/little

Add the correct amount of

 

medication in the medication

medication to the medication

 

tank?

tank.

 

 

 

 

Is the vaporiser head

Attach the vaporiser head

 

missing and not assembled

 

correctly.

 

correctly?

 

 

 

 

 

 

Is the nebuliser kit

Assemble the nebuliser kit

 

assembled correctly?

correctly.

 

 

 

No nebulisation or low

Is the nozzle blocked?

Make sure that the nozzle is

free of blockages.

nebulisation rate, when the

 

 

 

Make sure that the nebuliser

power is on.

Is the nebuliser kit tilted at a

 

kit is not tilted at an angle of

 

sharp angle?

more than 45 degrees.

 

 

 

 

 

 

Is the air tube connected

Make sure that the air tube is

 

correctly connected to the

 

correctly?

 

compressor and nebuliser.

 

 

 

 

 

 

Is the air tube folded or

Make sure that the air tube

 

damaged?

does not contain kinks.

 

 

 

 

Is the air tube blocked?

Make sure that the air tube is

 

free of blockages.

 

 

 

 

 

 

Is the air filter dirty?

Replace the air filter with a

 

new one.

 

 

The device is abnormally

Is the air filter cover attached

Attach the air filter cover

loud.

correctly?

correctly.

 

 

 

The device is very hot.

Are the ventilation slots

Do not block the ventilation

blocked?

slots.

 

 

 

 

Notes:

If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable.

Return the unit to an authorized OMRON retail outlet or distributor.

The device should be serviced once a year by an approved OMRON service centre.

22

Image 22
Contents Model NE-C29-E Contents Introduction Intended useImportant safety instructions UsageImportant safety instructions General Safety Precautions Save these instructions for future referenceKnow your unit Rear ViewAccessories Guide to using the Comp Air Pro Preparing the nebuliser for use Capacity of the medication tank is To 7 ml Attaching the adult mask PVC Preparing the nebuliser for use Taking medication To stop nebulisationUsing the mouthpiece Power Plug Cleaning and daily disinfecting Disinfecting the nebulising parts Compressor is not waterproof Handling the Vaporiser Head Maintenance and storage Changing the air filter Troubleshooting Symptom Cause RemedyTechnical data ContentsMmad = Mass Median Aerodynamic Diameter = TypeBB Technical data Optional parts and accessories Made in China Modèle NE-C29-E