8. Troubleshooting
Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions.
Symptom | Cause | Remedy | |
|
|
| |
Nothing happens when the | Is the power plug plugged | Check that the plug is | |
inserted in an electrical | |||
power switch is pressed. | into an electrical outlet? | outlet. Unplug then reinsert | |
|
| the plug if necessary. | |
|
|
| |
| Is there medication in the | Fill the medication tank. | |
| medication tank? | ||
|
| ||
|
|
| |
| Is there too much/little | Add the correct amount of | |
| medication in the medication | medication to the medication | |
| tank? | tank. | |
|
|
| |
| Is the vaporiser head | Attach the vaporiser head | |
| missing and not assembled | ||
| correctly. | ||
| correctly? | ||
|
| ||
|
|
| |
| Is the nebuliser kit | Assemble the nebuliser kit | |
| assembled correctly? | correctly. | |
|
|
| |
No nebulisation or low | Is the nozzle blocked? | Make sure that the nozzle is | |
free of blockages. | |||
nebulisation rate, when the |
|
| |
| Make sure that the nebuliser | ||
power is on. | Is the nebuliser kit tilted at a | ||
| kit is not tilted at an angle of | ||
| sharp angle? | more than 45 degrees. | |
|
| ||
|
|
| |
| Is the air tube connected | Make sure that the air tube is | |
| correctly connected to the | ||
| correctly? | ||
| compressor and nebuliser. | ||
|
| ||
|
|
| |
| Is the air tube folded or | Make sure that the air tube | |
| damaged? | does not contain kinks. | |
|
|
| |
| Is the air tube blocked? | Make sure that the air tube is | |
| free of blockages. | ||
|
| ||
|
|
| |
| Is the air filter dirty? | Replace the air filter with a | |
| new one. | ||
|
| ||
The device is abnormally | Is the air filter cover attached | Attach the air filter cover | |
loud. | correctly? | correctly. | |
|
|
| |
The device is very hot. | Are the ventilation slots | Do not block the ventilation | |
blocked? | slots. | ||
| |||
|
|
|
Notes:
•If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable.
Return the unit to an authorized OMRON retail outlet or distributor.
•The device should be serviced once a year by an approved OMRON service centre.