4.5If after completing step 4.4, the On Target LED (1) blinks erratically or not at all, rotate the Sensitivity Adjust knob (7) slowly from 1 to 9 until the LED blinks regularly or is on solid.

4.6If after completing steps 4.1-4.5, the On Target LED (1) blinks erratically or not at all:

4.6.1Move the sensor closer to the target and ensure it is perpendicular to the target. Repeat steps 4.1-4.5.

4.6.2Increase the contrast/reflectivity of the target. Repeat steps 4.1-4.5.

4.7Once the On Target signal is obtained, use the Polarity Switch

(5) to select a compatible output pulse for your application.

Weight: 10.6 oz (300.5 grams) excluding cable and charger

Compliance: CE 93/68/EEC EMC directive (EN 61326:2001)

IEC 61010-1 2001, IEC 60825-1 2001, CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001

N.I.S.T. Calibration (RS232 output only)

Mounting: ¼-20 UNC Bushing included (Tripod Mount)

Installation Environment:

Installation Category II

per IEC 664

Pollution Degree Level II

per IEC 61010-1

5.0 CONNECTION DETAIL

The unit has an input power jack for DC power or recharging the batteries. Use only the PR Universal charger supplied.

The unit also has a five-pin DIN output socket for Pulse outputs and RS232. An 8-foot [2.5 m] cable with a DIN plug and BNC connector is supplied. The BNC plug is connected to the SO output. A variety of optional cables are available including: tinned wire termination, 1/8 inch [3.5 mm] mono plug, or DB9 for RS-232.

Pulse/

RS232

Output

DC Power/Charger

9V 3VA MAX

Use PR Universal ONLY.

Temperature: 32 to 104 °F (0 to +40 °C) operating 14 to 158 °F (-10 to +70 °C) storage

Humidity: Maximum relative humidity of 80% for temperatures up to 88 °F (31 °C) decreasing linearly to 50% relative humidity at 104 °F (40 °C)

Specifications subject to change without notice.

2.0 OVERVIEW

The Smart Laser Sensor (SLS) is a self-contained unit intended to be used to make non-contact speed measurements from rotating targets at distances up to 65 feet (19.8 m) or to provide non-contact reference points to balancing equipment. Refer to Section 6.0 for an overview of operation. The unit will output one (TTL Compatible) pulse per revolution and has the ability to compute the RPM internally and output the ASCII values to any equipment capable of receiving an RS232 input. For best performance use reflective tape for the target.

Figure 4 Connection Details

7

4

Image 7
Contents 1071-4859-111 CE Declaration of ConformityMonarch Instrument Monarch InstrumentExamples 1234.56cr 123456.cr 0cr = space character Safeguards and PrecautionsAccessories ChargingRS232 Aiming the LaserLaser is on when this indicator is illuminated Smart Laser Sensor SpecificationsAdvanced Feature Target Polarity Laser SpecificationsOperation Auto ModeOutput Connections ConnectorDimensions RS232Per IEC Installation Environment Installation CategoryConnection Detail OverviewQuick Start Guide Indicators and Switches