Agilent Technologies 60502B manual Specifications and Supplemental Characteristics

Page 7

 

Table 60502-1 Specifications and Supplemental Characteristics (continued)

Current Slew Rate:*

 

 

 

Rate # 60 A Range Step

6 A Range Step

Transition Time

1

1 A/ms

0.1

A/s

8.0 ms

2

2.5 A/ms

0.25 A/s

3.2 ms

3

5 A/ms

0.5

A/s

1.6 ms

4

10 A/ms

1 A/ms

800 μs

5

25 A/ms

2.5

A/ms

320 μs

6

50 A/ms

5 A/ms

160 μs

7

0.1 A/μs

10 A/ms

80 μs

8

0.25 A/μs

25 A/ms

32 μs

9

0.5 A/μs

50 A/ms

16 μs

10

1 A/μs

0.1

A/μs

12 μs

11

2.5 A/μs

0.25 A/μs

12 μs

12

5 A/μs

0.5

A/μs

12 μs

 

*AC performance specified from 3 to 60 V.

Voltage Slew Rate:

 

 

Rate # Voltage Range Step

Transition Time*

1

1 V/ms

8.0 ms

2

2.5 V/ms

3.2 ms

3

5 V/ms

1.6 ms

4

10 V/ms

800 μs

5

25 V/ms

320 μs

6

50 V/ms

160 μs

7

0.1 V/μs

85

μs

8

0.25 V/μs

85

μS

9

0.5 V/μs

85

μS

*Transition time based on low capacitance current source.

Resistance Slew Rate (1 Ω range): Uses the value programmed for voltage slew rate.

Resistance Slew Rate (1 k and 10 kΩ ranges): Uses the value programmed for current slew rate.

Transient Current Overshoot (When programmed from 0A):

 

Range

Transient Current Level

Current Slew Rate

Overshoot*

60 A

6-60 A

All slew rates

0

 

3 A

1 A/μs to 5 A/μs

1%

 

3 A

1 A/μs to 0.5 A/μs

0

6 A

6 A

All slew rates

0

 

3 A

0.25 A/μs to 0.5 A/μs

1%

 

3 A

0.1 A/ms to 0.1 A/μs

0

*Overshoot may be higher during the first five seconds of programming if unit has been operating at full current. Overshoot values assume a total inductance of lμH, or less, in the load leads connected to the D.U.T.

5

Image 7
Contents 3118A-00101 and Above Agilent Part No Microfiche No June For Modules with Serial NumbersEMC Watt Module Operating Characteristics Derated Current Detail SpecificationsTransient Voltage Level 0 to 60 Resolution Frequency Resolution Frequency Accuracy Duty Cycle RangeDuty Cycle Resolution Duty Cycle Accuracy Pulsed Mode Pulse WidthSupplemental Characteristics Voltage Slew Rate Rate # Voltage Range Step Transition Time Specifications and Supplemental CharacteristicsTransient Current Overshoot When programmed from 0A Current Slew Rate Rate # 60 a Range Step Transition TimeConstant Resistance Weight Function Constant CurrentReverse Current Capacity Range of ValuesFreq value Tranfreq value Shift Dcycle value Trandcyc value Factory Default Settings FunctionVariables Value Power Supply SettingsCalibration Information Ranges and Calibration Points VariablesAgilent Sales and Support Office
Related manuals
Manual 42 pages 10.26 Kb

60502B specifications

Agilent Technologies 60502B is a high-performance signal generator designed for a wide range of applications, including aerospace, telecommunications, and research and development. As part of Agilent's advanced suite of test and measurement solutions, the 60502B provides users with the capability to generate high-quality signals with exceptional accuracy and stability.

One of the standout features of the Agilent 60502B is its frequency range, which spans from 9 kHz to 6 GHz. This broad range enables engineers and technicians to work with various modulation formats and signal types, making it highly versatile for testing and development. The generator supports continuous wave (CW) signals as well as complex modulated signals, accommodating both narrowband and broadband applications.

The 60502B employs advanced direct digital synthesis (DDS) technology, ensuring accurate frequency generation with low phase noise. This technology is essential for applications that require high precision and minimal signal distortion, such as radar and communication systems. The generator’s phase noise performance is exceptional, which enhances its capability for demanding applications requiring signal integrity.

Another key characteristic of the Agilent 60502B is its output power handling. It can deliver output levels ranging from -130 dBm up to +10 dBm, enabling users to drive various loads and achieve optimal testing conditions. This flexibility is particularly beneficial in laboratory environments where different power levels are needed for different devices under test.

The user interface of the 60502B is designed for efficiency and ease of use. It incorporates a large display that provides real-time feedback on settings and signal characteristics, facilitating quick adjustments during testing. Furthermore, the generator is equipped with USB and GPIB interfaces, enabling seamless integration into automated test systems and easy connectivity to computers for remote control and data logging.

In addition to its robust hardware features, the Agilent 60502B supports a variety of modulation schemes, including AM, FM, PM, and pulse modulation. This capability allows users to simulate real-world signals and assess the performance of receivers and other components under various conditions.

Overall, the Agilent Technologies 60502B signal generator stands out in terms of performance, versatility, and user-friendliness. Its wide frequency range, advanced DDS technology, significant power output capabilities, and support for multiple modulation types make it an indispensable tool for engineers across various industries engaged in testing and development activities. Whether for research projects or routine maintenance tasks, the 60502B is built to deliver reliable and precise results.