HP TM 11-6625-2779-14&P Performance Checks, Level Accuracy Checks, Function CARRIER, 75 Unbal

Page 36

Section V

Model 3555B

Figure 5-1. Balanced BNC to 310 Plug

function, remove or clip the shorting bar from across A1 R17 (see Figure 7-2).

b.Remove the 150 BAL decal from the small envelope supplied with the set and stick it over the existing 135 BAL decal.

c.Adjust the 150 function as described in Paragraph 5-20in this manual.

5-7. PERFORMANCE CHECKS.

5-8. The performance checks presented in this section are in-cabinet checks designed to compare the Model 3555B with its published specifications. These checks can be used for incoming inspection, periodic maintenance checks and to verify performance after adjustment or repair. A performance check test card appears at the end of this section which can be used to record the specification performance of your set.

5-9. LEVEL ACCURACY CHECKS.

a.Connect only the 654A and 3555B as shown in Figure 5-2and set the 3555B controls as follows:

FUNCTION

CARRIER, 75 UNBAL

INPUT

TMS, TERM

RANGE

+10dBm

b. Set the 654A frequency to 20kHz, IMPEDANCE to 75 UNBAL and adjust the output level for +10dBm. If the calibration of the 654A is questionable, first connect the output of the 654A through a 75 ohm termination, directly to the input of the 3440A/3445A (3555B not connected) and measure the voltage. This level should be 866mV rms. If it is not, adjust the 654A amplitude control until it is and note the 654A meter indication for future reference. Now that the 654A calibration has been verified, disconnect

Figure 5-2. Level Accuracy Check

5-2

Image 36
Contents HEADQUARTERS, Department of the Army Reporting Errors and Recommending Improvements TM 11-6625-2779-14&PTechnical Manual Headquarters Department of the Army NSNReplaceable Parts List of Illustrations List of TablesIii Maintenance Forms, Records, and Reports Section IntroductionScope Administrative Storage Destruction of Army Electronics MaterielReporting Equipment Improvement Recommendations EIR Items Comprising an Operable Equipment Specifications Voice Frequency Level Measurements 20Hz to 20kHzSection Section General Information IntroductionAccessory Equipment Supplied Accessory Equipment SuppliedInstrument Identification BAL ModificationSection Installation Inspection Warranty ExceptionPower Requirements BatteryInstallation and Removal of Battery Cover RemovalControls Connectors Indicators BatterySection Operating InstructionsFront Panel Controls, Indicators, and Connectors Front, Side and Rear Pane Side Panel Controls and Connectors Level and Noise Measurements Level MeasurementsPage Noise Measurement Recorder Compatibility ApplicationsRecorder Compatibility Chart Identifying Noise Characteristics Transmission Loss MeasurementsCrosstalk Measurements Crosstalk Correction FactorMeasurements in DBC Reinstall the set in its case 0dBmInput jacks are connected in parallel. If a 600 ohm 135 BALStep Procedure Level Measurement Noise Metallic MeasurementsNoise-to-Ground Measurements Balance Measurement Recorder CalibrationExample Transmission Loss MeasurementBlock Diagram Description Section Theory of Operation IntroductionDetailed Circuit Description Simplified Dial BAT FunctionInput Amplifier A3. Schematic No Range Attenuator A2Setting Attenuation FILTERS. Schematic NoRange Attenuator Pads UsedKHz Flat and Program Weighting Curves C-MSG and 1SkHz Flat Weighting Curves Meter AMPLIFIER. Schematic No DETECTOR. Schematic NoPower Supply and Series Regulator Schematic No Section Maintenance Introouction Instrument Recommended Type Required Characteristics ModelFactory Selected Values BAL ConversionPerformance Checks Level Accuracy ChecksFunction CARRIER, 75 Unbal TMS, TermUnbal Carrier Accuracy Check RangeFrequency 30Hz to 1MHz 100Hz to 600kHz 1MHz to 3MHzReturn Loss Check 20Hz to 20kHz 40Hz to 15kHz 100Hz to 20kHz 100Hz to 15kHzReturn Loss Test Set-Up Filter Response Checks Filter Response ChecksMSG dBm 3kHz Flat dBm 15kHz Flat dBm Program dBm Input Balance Bridging LossTMS Brdg Input TMS BrdgAdjustment and Calibration Procedure Power Supply CheckUnbal Calibration Attenuator CalibrationFrequency Response Adjustment Common Mode AdjustmentBalance Check Assembly Removal Troubleshooting ProceduresFilter Calibration Troubleshooting Tree Front Panel Trouble Analysis Dial BATFront Panel Troubleshooting Function TroubleshootingRange Troubleshooting Front Panel Trouble Analysis ContdTroubleshooting Input Amplifier Filter TroubleshootingTroubleshooting the Meter Amplifier and Detector Defective FunctionFunction Switch Resistance Values Function Input Jacks DIAL/AC MON JacksTip to Ring Tip to Ground Ring to Ground Brdg Term11. Factory Selected Values 10. Resistance ChecksRange dBm Pin 1 to Pin 2 to Performance Check Test Card Description CheckCarrier 75 Unbal Level Accuracy Check Carrier 135 BAL Level Accuracy CheckPerformance Check Test Card Contd Carrier 600 BAL Level Accuracy CheckReturn Loss Check Filter Response Checks MSGProgram Bridging Loss CheckSection Replaceable Parts Introduction Ordering InformationReplaceable Parts 03555-66507 PC Board Assy functionReplaceable Parts Contd 03555-66508 Board Assy amplifierA3 Contd 03555-66506 PC Board Assy filter A4 Contd Chassis Mounted Components Chassis Mounted Components Contd MiscellaneousMiscellaneous Contd Replacement National Part Stock Number Fscm Part Number National Stock Number Cross Reference IndexFscm National Stock NumberPage Model 3555B Appendix a Code List of Manufacturers Circuit Diagrams Functional Block DiagramSchematic Diagrams Reference DesignationsSection Model 3555B Functional Block Diagram 7-4A1 Function Assembly Schematic and Component Location 7-67-8 A4 Filter Schematic and Component Location 7-1011/7-12 Appendix a ReferencesMaintenance Allocation Section I. Introduction General Maintenance FunctionAppendix B Column EntriesTool and Test Equipment Requiremen ts Sec Remarks SecAudio Level Meter TA-885/U TM 11-7725-2779-14&P REF Code NumberAppendix C Manual Backdating ChangesTransmission and Noise Measuring SET Delete CR6 across M1 Government Printing Office 1980-603-128/66General, United States Army Chief of StaffMajor General, United States Army Adjutant General Active ArmyPage PIN