Bacharach 24-9448 manual Memory, Ending the Data Logging Process

Page 49
Where: “yy” is the number of log entries stored during the last logging session.
yy Entries Stored
CO2 T-STKT-AIR EA CO(3)
EFF
CO
O2
LOG NATURAL GAS
F2
F1
F3
82.6 %
9.5 %
374 °F
68.0 °F
21.3 %
13 ppm
12 ppm
4.0 %

Operation

4.12.4Ending the Data Logging Process

Data logging will stop and the pump will turn OFF after the prescribed duration, or after all 500 memory locations are filled.

To exit the Combustion Test LOG screen, press the RUN/HOLD button twice to display the Com- bustion Test HOLD screen.

Note the following:

• To end the logging process at any time, press the ESC button.

• At the end of the logging period, or if the ESC button was pressed, the total number of log entries that were stored during that period appear at the bottom of the screen.

If the logging process was stopped by pressing ESC, it cannot be re- sumed without first turning logging back ON per Section 4.12.1.

4.13 Memory

There are two memory banks, each containing 500 memory locations. The first bank is used to store combustion test data as described in Section 4.7., while the second bank is used to store logged combustion test data as described in Section 4.12. Each bank is independent of each other, and can- not share data or be combined.

Individual memory locations in each memory bank can be recalled for viewing on the display or printed (refer to Sections 4.13.1 & 4.13.2), or the entire contents of each memory bank can be individually downloaded to a computer and viewed in a spreadsheet program for analysis (refer to Sec- tion 4.14).

TIP: When displaying the contents of either memory bank, the operator can quickly page through the screens by press- ing the PAGE– (F1) and PAGE+ (F3) buttons. Or move to the first or last memory location by pressing the buttons, respectively.

Instruction 24-9448

4-15

Image 49
Contents Portable Combustion Analyzer Warranty Contents Operation Maintenance Contents Introduction PCA 2 General Description Calculations Sales Combo & Model ConfigurationsCalculations Features & Benefits Operational Overview Probe Connections Gas, Pressure, T-Stack Connector DescriptionsAIR Primary Air Thermocouple PCA 2 Components 4 ∆P Differential Pressure Power AC AdapterUSB Computer Interface IrDA Printer InterfaceFront Panel Buttons Introduction PCA 2 Directly Measures and Displays SpecificationsNormal Operating Conditions Warm-Up Time Specifications Power Initial Setup ScopeInstalling or Replacing Batteries Using the AC Power Adapter Connecting the Probe and Hose AssemblyConnecting the Probe and Hose Assembly to the PCA Fuel Selection Operating ParametersTemp Units Menu Temperature Units SelectionPressure Units Menu Pressure Units SelectionPollution Units Menu Pollution Units SelectionDate Setup Menu Time Setup11 O2 Reference Setup Print Pressure Selection Zoom-Display Selection O2 4.0 % CO 12 ppm EFF 82.6%Battery Charger Selection Button Sound Logging SelectionOperation Operating Tips Turning on the Analyzer and Warm Up Menu F2 button Selecting a FuelSampling Point 80% Efficiency Fan Assist or Power VentedPerforming a Combustion Test EFFList of Combustion Test Data TIP The pressure units of measure is selected per Section Making a Draft / Pressure MeasurementSaving Test Data Emptying the Water Trap Ending a Combustion TestTurning OFF the Analyzer & Purging Data Logging Use the buttons to highlight YES turn logging on Turning on Data LoggingSetting the Logging Interval and Duration Starting the Data Logging Process Ending the Data Logging Process MemoryRecalling Combustion Test Data Recalling Logged Test Data Clearing Memory Downloading Stored Data to a Computer PCA 2 Data Recovery Program Installation Operation Computer to PCA 2 Connection Computer to PCA 2 Connection USB Device Driver InstallationAt the Hardware Installation window, click Continue Anyway Recovering Data Log / Data File button Additional­ locations that contain sensor calibration data Unplugging the USB Cable Importing Saved Data Into a Spreadsheet Downloaded Data Fields Printing Test Data Printer Alignment & Sample Printout Operation Smart Sensors CalibrationStarting a Calibration Benefits of Smart SensorsPressure Sensor Calibration ProcedureMaterial Required Pressure Sensor Calibration Equipment Stack Calibration TS-Zero ProcedureTS-Span Procedure Air Calibration TA-Zero ProcedureTA-Span Procedure CO-LO Sensor Calibration Turn OFF the regulator and remove the gas cylinder SO2 Sensor Calibration No Sensor Calibration NO2 Sensor Calibration CO-HI Sensor Calibration Gas Sensor Calibration Equipment Maintenance Removing Rear Case PCA 2 DisassemblySensor Positions Tubing Connections Wiring Emptying the Water Trap Chamber Water Trap / Filter MaintenanceReplacing the Filter Element Smart Sensor Replacement Sensor Only Replacement Maintenance Bias battery removed for Stabilization time Nitric Oxide Sensor Battery ReplacementTools Required Thermocouple ReplacementThermocouple Installation and Wiring Parts & Tools Required Pump ReplacementEquipment Required Cleaning the ProbeMaintenance Instruction Error Symbols Troubleshooting Analyzer RepairError Messages Displayed After Warm-Up Diagnostics and Status Screens Troubleshooting Description Parts & Service Replacement PartsDescription Parts 1 Parts 2 Standard Accessories AccessoriesOptional Accessories United States Service CentersPage Headquarters Hunt Valley Circle, New Kensington, PA

24-9448 specifications

The Bacharach 24-9448 is a sophisticated instrument designed for detecting and measuring hazardous gases, particularly in environmental and industrial settings. Renowned for its accuracy and reliability, this portable gas analyzer serves a crucial role in ensuring safety and compliance in various applications, including HVAC, refrigeration, and environmental monitoring.

One of the standout features of the 24-9448 is its advanced sensor technology. The device utilizes electrochemical sensors that provide real-time readings of specific gas concentrations, including carbon monoxide, carbon dioxide, and other toxic gases. These sensors are designed to be sensitive and selective, ensuring they deliver precise measurements without interference from other substances in the environment. This is particularly important in industrial settings where multiple gases may be present.

The Bacharach 24-9448 also boasts a user-friendly interface with a clear display that makes it easy for operators to interpret results quickly. The unit features a simple menu navigation system, allowing users to access different functions without hassle. This design consideration enhances usability, especially in high-pressure situations where every second counts.

Portability is another key characteristic of the 24-9448. Weighing in at just a few pounds, it is easy to transport and operate in various locations, whether it be on a job site or in an environmental monitoring scenario. The robust housing is designed to withstand the rigors of fieldwork, ensuring longevity and durability despite challenging conditions.

Moreover, the device includes data logging capabilities, enabling users to store and later review measurements. This feature is essential for compliance documentation and for analyzing trends over time. Users can download data for further analysis, aiding in thorough reporting and decision-making processes.

In addition to its core capabilities, the Bacharach 24-9448 offers flexibility with adjustable calibration settings, accommodating different gas types and concentration ranges. This adaptability ensures it can meet the specific needs of diverse applications, making it a versatile tool in the field.

Overall, the Bacharach 24-9448 is an essential instrument for professionals in various sectors. Its combination of advanced sensor technology, user-friendly design, portability, and data logging capabilities solidifies its position as a leading choice for gas detection and measurement. Whether for routine checks or emergency responses, this analyzer provides the reliability that ensures safety in potentially hazardous environments.