Fluke 61 user manual Specifications, Accuracy

Page 2

 

 

Specifications

 

 

Temperature range

-18 to 275°C (0 to 525°F)

 

 

 

 

 

 

Display Resolution

0.2°C or 0.5°F

 

 

 

 

 

 

Accuracy

For targets at:

 

 

(assumes ambient operating

-1 to 275°C (30 to 525°F) ±2% of reading

 

 

temperature of 23°C [73°F])

or ±2°C (±3.5°F), whichever is greater

 

 

 

-18 to -1°C (0 to 30°F) ±3°C (±5°F)

 

 

 

 

 

 

Temperature Coefficient

0.2K per °C or 0.2% per °C, whichever is greater

 

 

 

 

 

 

Repeatability

± 2% of reading, or ± 2°C (±3.5°F)

 

 

 

whichever is greater

 

 

 

 

 

 

Response time

500 mSec, 95% response

 

 

 

 

 

 

Spectral response

7–18 µm

 

 

 

 

 

 

Emissivity

pre-set 0.95

 

 

 

 

 

 

Ambient operating range

0 to 50°C (32 to 120°F)

 

 

 

 

 

 

Relative humidity

10–90% RH noncondensing, @ up to 50°C (120°F)

 

 

 

 

 

 

Storage temperature

-20° to 60°C (-4° to 140°F) without battery

 

 

 

 

 

 

Weight / Dimensions

227 g (0.5 lb); 184 x 45 x 38 mm

 

 

 

(7.25 x 1.75 x 1.5 in) without holster

 

 

 

341 g (0.75lb); 190 x 51 x 41 mm

 

 

 

(7.5 x 2.0 x 1.6 in) with holster

 

 

 

 

 

 

Power

9V Alkaline or NiCd battery

 

 

 

 

 

 

Battery life (Alkaline)

12 hrs

 

 

 

 

 

 

Distance to Spot Size

8:1

 

Specifications subject to change without notice.

Image 2
Contents Fluke-61 Specifications AccuracyBattery Cover Introdução Atenção modelos com mira laserCuidados Como funcionaComo operar a unidade Bateria e C/FComo Medir Temperaturas Com Precisão LembretesManutenção Identificação de problemasEmissividade Certificado ce Garantia Limitada E Limitação DE ResponsabilidadeNoncontact Thermometer Uses Contacting Fluke