1)Remove packaging materials, packaging bag and straps. Open the refrigerator door to let it air out. If there are dirt marks on the outside of the unit, use a soft sponge with a light detergent solution to wipe the unit clean. Then, rinse the unit with a clean, slightly damp towel.

2)Adjust the height of the unit. Turn the leveling leg screw counterclockwise to level the front end of the unit. The front casters should be off the ground.

3)Preparation of sensor bottle Before using the refrigerator, inspect the sensor bottle located at the back of the drawers and shelves to see if the liquid level is up to the 200 mark inside the bottle. If the bottle has no fluid or the fluid level is lower than the 200 mark, use the following steps to fill the bottles at the upper corner and the lower corner with solution of 10% glycerin concentration (or other equivalent liquid such as glycol solution).

a)Remove the top drawer, bottom drawer, shelves and baskets. b)Unscrew the fastener that positions the bottle, remove the bottle. c) Unscrew the bottle's cap that is used for the sensor mounting.

d)Pour a solution of 10% glycerin concentration into the bottle(s) to the 200 mark line. e)Mount the bottle back to the refrigerator liner wall.

f)Install the bottle back to its original position. Place the temperature sensor back inside the temperature bottle.

Operation

1)Supplementary power supply (9 volt DC).If the LED light is off and the recorder stops working, it would suggest that there might be issues associated with the main power supply. Please inspect whether there is a loss of power, a short circuit, or a loose connection. If there is no issue with the main power supply and the recorder starts working normally, please install the backup battery to the recorder. At this time, if the LED stops flashing, the supplementary power supply is normal. If the LED continues to flashes, it means that the power supply is under capacity requirement. It is necessary to replace the battery. The LED green light is off when there is sufficient power to the recorder.

2)Remove the supplementary power supply. Power LED light continues to flash. The recorder works normally. It suggests that the main power supply is sufficient while the backup battery lacks capacity. It is then necessary to replace the battery until the green LED lights up and stops flashing.

3.Changing of Recorder Chart Paper

When changing the chart paper for the recorder, please use the following steps. a.Locate the #3 button at the front of the recorder.

b.Press and hold the #3 button for about one second. The imprinting stylus of the recorder moves to the left side of the chart.

c.When the stylus is completely off the chart, unscrew the center nut for the paper and remove the old chart paper off the recorder. Place a new chart paper on the recorder. Carefully match the day and time line on the chart with the reference mark on the recorder panel (there is a small groove on the front of the recorder as shown in the figure).

d.Place the center nut over the chart paper and tighten it. Press the #3 button to move the stylus back to the chart to continue the temperature recording.

e.Inspect the contact between the pen and the paper chart. Adjust the arm if necessary to make contact. (Note, do not damage the pen and arm. It might be easier to remove the arm and pen to make a small bending adjustment and then install the pen back on the recorder. Repeat the process until the recorder arm and pen work properly.)

f.The accuracy of the temperature must be checked after changing the chart paper. The procedure is described as follows. Press and hold the #3 button until the pen moves to the outside of the chart paper. Then press the #3 button again; the pen should return to the chart paper area. When the pen reaches the outer most circle of the chart paper, it pauses for a moment. If the pen does not match the outer circle of the chart marks, you can use the #2 or the #1 button to move the pen to match the line. This must be done within 5 seconds after the pen pauses. If the calibration is incomplete, repeat the same process until the pen matches the outer circle line.

4.Calibration of Temperature Recorder

This recorder has been calibrated at the factory. It retains the temperature accuracy and calibration even when there is a power supply interruption. Should there be a need to recalibrate the recorder, please follow the procedures below.

a)Let the blood bank run continuously until it reaches a stable condition. The recorder has registered temperature for at least two hours.

b)Place a standard thermocouple in a 100 ml container filled with 10% glycerin solution. Place the container and the temperature recorder sensor in the refrigerator. Do not place the recorder sensor in contact with any liquid.

c)After four hours of soak time at the stable refrigerator temperature, the container temperature should match the refrigerator temperature. Then compare the thermocouple reading with the temperature recorder reading. If there is a discrepancy, use the #1 or the #2 button on the recorder to move the pen on the chart to reflect the correct temperature. Please note the pen moves five seconds after the button is released.

Note:that the recorder has been calibrated in the factory. Unless it is absolutely necessary, please do not change the setting.

Image 8
Contents HYC-610 Pharmaceutical RefrigeratorTable of Contents Safety Precautions Example of labelsAfter Sale Service Dear CustomerOperating Parameters SpecificationDescriptions of Refrigerator Parts Their Functions Maintenance Proper usage of the Refrigerator Location of RefrigeratorInstallation Changing of Recorder Chart PaperCalibration of Temperature Recorder OperationDefrost Cycle Temperature Recorder optionalControl Panel Solution Procedure to Test the Alarm SystemTrouble Shooting Guide

HYC-610 specifications

The Haier HYC-610 is a modern and innovative refrigerator that stands out in today’s competitive market for appliances. This model is designed not only to optimize storage but also to enhance energy efficiency, making it a top choice for environmentally conscious consumers.

One of the main features of the HYC-610 is its spacious interior. With a generous capacity, this refrigerator provides ample space for storing a variety of items, from fresh produce to frozen goods. The adjustable shelving system allows users to customize their storage according to their needs, making it easier to organize food items efficiently.

An important technology incorporated in the Haier HYC-610 is the advanced cooling system. This system ensures that temperature is maintained evenly throughout the compartments, which helps in keeping food fresh for longer periods. The refrigerator uses a multi-airflow system that circulates cold air evenly to all shelves, hence preventing hotspots and ensuring optimal cooling conditions at all times.

Another key characteristic of the HYC-610 is its energy efficiency. It comes equipped with an A++ energy rating, which signifies its ability to consume less power compared to conventional refrigerators. This not only helps in reducing electricity bills but also contributes to lower carbon emissions, making it an eco-friendly choice for consumers.

The HYC-610 also features a user-friendly control panel that makes it easy to adjust settings and temperatures according to personal preferences. Additionally, the quick freeze function is particularly useful for rapidly cooling down items, preserving their nutrients and flavor.

Hygiene is taken into consideration with the refrigerator’s antibacterial coating in critical areas, ensuring that the interior remains clean and free from harmful bacteria. This focus on health aligns with contemporary consumer trends that prioritize safety and cleanliness in food storage.

In terms of design, the Haier HYC-610 exhibits a sleek and modern aesthetic, making it a stylish addition to any kitchen. The stainless-steel finish not only enhances appearance but is also easy to clean, promising longevity and durability.

Overall, the Haier HYC-610 combines functionality, efficiency, and style, making it a standout choice for consumers seeking reliable refrigeration solutions. With its innovative technologies and thoughtful design, the HYC-610 represents a blend of modern convenience and environmental responsibility.