Emerson Process Management 2090, 2088 manual General Considerations, Environmental Considerations

Page 4

Rosemount 2088 and 2090

Reference Manual

00809-0100-4690, Rev EA March 2007

GENERAL CONSIDERATIONS

The accuracy of the pressure measurement depends on proper installation of the transmitter and impulse piping. The piping between the process and transmitter must accurately transmit pressure to the transmitter. Mount the transmitter close to the process and use a minimum of impulse piping to achieve the best accuracy. Keep in mind, however, the need for convenient access, safety of personnel, practical field calibration, and a suitable transmitter environment. In general, install the transmitter to minimize vibration, shock, and temperature fluctuations.

Do not apply torque directly to the electronics housing. Rotation between the electronics housing and the process connection can damage the electronics. To avoid damage, apply torque only to the hex-shaped process connection. Use a backup wrench on Process Connection when installing an adapter.

Electronics

Housing

Process Connection

ENVIRONMENTAL CONSIDERATIONS

Temperature

Moisture and Corrosives

Hazardous Locations Installations

Mount the transmitter in a manner that minimizes variations in ambient temperature.

The transmitter is designed to resist attack by moisture and corrosives. The electronics module is fully encapsulated and mounted in a compartment that is sealed from the power-side conduit entries. O-ring seals protect both compartments when the covers are installed.

In humid environments, it is possible for moisture to accumulate in the conduit lines and reach the terminal compartment of the transmitter housing. To prevent moisture from entering the terminal compartment, mount the transmitter at a high point in the conduit run, if possible. Also, remove the terminal compartment cover periodically and inspect the terminals for moisture and corrosion.

Rosemount 2088, 2090P, and 2090F transmitters are designed with explosion-proof electronics enclosures and circuitry that complies with intrinsic safety requirements and non-incendive operation. Individual transmitters are clearly tagged with approvals. Refer to Section 5: Specifications and Reference Data for a complete list of available approvals. To maintain certified ratings for installed transmitters, install with applicable installation codes and approval drawings.

NOTE

Once a device labeled with multiple approval types is installed, it should not be reinstalled using any other approval types. Permanently mark the approval label to distinguish it from unused approval types.

3-4

Image 4
Contents Overview Safety Messages Reference ManualRosemount 2088 Installation Flowchart General Considerations Environmental ConsiderationsMechanical Considerations MountingImpulse Piping Access Requirements Rosemount 2090P 1.5-inRosemount 2090P Compatible with 1-in. PMC Process Connection Rosemount 2090P Detail of Weld Spud Rosemount 2090F Power SupplyElectrical Considerations Field Wiring 10.5Failure Mode Failure Mode and Security JumpersJumper Locations 11. Transmitter Alarm Security Jumper Locations Transmitter SecurityZero and Span Adjustments Rerange ProcedureRosemount 2088

2088, 2090 specifications

Emerson Process Management, a global leader in process automation and control solutions, has developed a series of advanced controllers including the 2090 and 2088 models. These controllers are integral to modern industrial applications, enhancing operational efficiencies and ensuring precise control over processes.

The Emerson 2090 controller is renowned for its high performance and reliability. It supports both continuous and batch processes, making it versatile for various industrial environments. One of its main features is its modular architecture, which allows for easy scalability and customization according to specific operational needs. The controller is equipped with advanced communication capabilities, supporting multiple protocols such as Modbus, Ethernet/IP, and Profibus, ensuring seamless integration with existing systems.

The 2090 also boasts sophisticated data processing abilities. It employs predictive analytics, enabling users to forecast potential issues and optimize process performance before disturbances occur. With its intuitive user interface and graphical programming environment, operators can design control strategies easily, significantly reducing programming time and enhancing user experience.

On the other hand, the Emerson 2088 controller offers a powerful yet cost-effective solution for smaller-scale applications. Designed for simplicity and ease of use, it contains essential control features while maintaining flexibility. It supports a variety of input and output modules, allowing for extensive configuration options to meet specific requirements. The 2088 is particularly noted for its reliability in extreme conditions, making it an ideal choice for demanding environments.

Both the 2090 and 2088 controllers utilize Emerson’s latest technologies, including advanced algorithms for PID control, and they support redundancy options for critical applications. The emphasis on cybersecurity in both models further ensures safe operation, protecting sensitive data and control systems from potential threats.

These controllers exemplify Emerson's commitment to innovation, efficiency, and reliability in process management. Their robust design and advanced features cater to a wide range of industries, including oil and gas, chemical processing, and water treatment. As industry standards evolve, Emerson continues to lead the charge by integrating cutting-edge technologies into their controller solutions to meet the ever-changing demands of process automation.