3Com 61-362 technical manual Replacing Fuse refer to Figure

Page 7

Page 5

Replacing Fuse (refer to Figure 3)

1.Disconnect the test leads and turn the range switch to “OFF”.

2.Remove the three screws holding the bottom case cover

3Us a digital multimeter in low resistance {ohms} mode to check the two fuses mA input, 0.5A / 250V fast acting fuse.

Amp input, 10A / 600V fast acting fuse

4.Remove the defective fuse with the recommended fuse

mA with : 0.5A/250V fast acting fuse Type LA-3895 is recommended.

Amp with: 10A/600V fast acting fuse, Type LA-3897 is recommended

5. Install bottom cover and secure with screws.

Form number 61362

rev 1 May 2002

Image 7
Contents Model Ideal Industries INC Technical ManualTable of Contents Specific precautions SafetyOverload Protection General specificationsRanges and Accuracy Specification 61-362 Function Setting AC ConverterPerformance Verification Preparation Performance VerificaitonsCalibration Preparation Calibration ProcedureBattery Replacement refer to Figure CalibrationReplacing Fuse refer to Figure