ResMed VS III user manual Switching programs during ventilation

Page 17

4.2 Switching programs (during ventilation)

 

 

 

27/09/2008

17:50

 

 

 

 

 

 

 

 

 

 

 

9

 

 

 

 

cmH2O

Prog1: PS.Vs

60

 

 

50

 

 

 

 

 

40

 

 

Prog2: PS.Vt

 

 

30

 

 

 

 

20

 

 

 

 

 

10

0Stop

!

Confirmation

 

9

Program change?

cmH2O

60

 

 

50

40No

30

Yes

20

10

 

0

Confirm

 

 

LP

27/09/2008 17:50

 

 

 

 

 

 

 

11

 

 

cmH2O

Prog2: PS.Vt

60

 

50

 

 

 

40

 

Prog1: PS.Vs

30

 

20

 

 

 

10

0Stop

1

2

While the device is delivering ventilation under “Prog1”, press .

On the program change confirmation

screen, use to select “Yes”, then press to confirm.

The device is now delivering ventilation under “Prog2”.

 

27/09/2008 17:50

 

 

27/09/2008 17:50

LP

 

 

LP

 

11

 

11

11

 

11

cmH2O

Prog2: PS.Vt

cmH2O

Prog2: PS.Vt

60

60

50

Prog1: PS.Vs

cmH2O

50

Prog1: PS.Vs

cmH2O

30

30

40

 

 

40

 

 

20

 

 

20

 

 

10

 

 

10

 

 

0

Stop

 

0

Stop

 

With “Mouthpiece/Speaking” With “Mouthpiece/

option deactivated

Speaking” option activated

LP Symbol displayed when the “Mouthpiece/Speaking” option has been selected for Prog 2 (in valve ventilation modes only).

Figure 13: Switching programs during ventilation

Note: Once the “Mouthpiece/Speaking” option is activated by your physician, you are able to speak without triggering the low pressure alarm during expiration.

Ensure that a caregiver is able to supervise the use of the device while this option is activated.

How to use

13

Image 17
Contents English Respiratory Care solutions Making quality of care easy Contents Page User/Owner responsibility IntroductionDefinitions Purpose of your ventilation device Medical informationYour device Description of the deviceComponents Air outlet Your deviceFront view LCD screen Rear viewDevice control panel Page Screen appearance while device operating Connection proceduresRespiratory circuit Connecting the device to the mains power supply Connecting to a power supplyOption 1 Single circuit Connecting the respiratory circuitAir return Option 3 Double circuitExample of a fully-assembled system Connecting the maskStarting ventilation How to useProgram ProgramsSwitching programs during ventilation Switching programs during ventilationEvent and Alarm Log Ventilation settings and alarm thresholdsViewing data during ventilation MeasurementsPS.Vs Setting the date and time while ventilation is StoppedProgram Programs Turning off the deviceStopping ventilation While ventilation is During ventilation Stopped Cleaning and maintenanceShutting down the device Clean the mask in warm Immediately TroubleshootingWithout delay and check Weight without external power supply Power supply Technical specificationsDimensions Transport Appendix1500h Symbols displayed on the screenPage Index Page Respiratory Care solutions Making quality of care easy 0123
Related manuals
Manual 6 pages 14.5 Kb

VS III specifications

ResMed VS III is a cutting-edge ventilator designed for both hospital and home use, specifically tailored to meet the needs of patients with respiratory challenges. It incorporates advanced technologies to provide high-quality respiratory support while ensuring comfort and versatility.

One of the main features of the ResMed VS III is its advanced flow and pressure monitoring capabilities. This ventilator can adjust to the patient's requirements in real-time, ensuring optimal therapy delivery. By utilizing intelligent algorithms, the device is capable of distinguishing between different breathing patterns, allowing it to automatically adjust settings to meet individual needs. This adaptability not only enhances patient compliance but also promotes better clinical outcomes.

The device also boasts a user-friendly interface, making it easier for healthcare providers to initiate and monitor treatment. The LCD display presents comprehensive data, allowing clinicians to quickly assess the patient's status and make necessary adjustments. This streamlined interface minimizes the learning curve for new users and facilitates efficient patient management.

Another significant advancement in the ResMed VS III is its built-in connectivity features. With integrated wireless capabilities, the device allows for remote monitoring and updates. Clinicians can easily access patient data, enabling them to make informed decisions without requiring the patient to visit the clinic. This is particularly beneficial for managing patients in home care settings, where mobility may be limited.

The ResMed VS III also emphasizes patient comfort through its selection of ventilation modes. The device supports a variety of modes, including pressure support, volume control, and spontaneous breathing, allowing for individualized therapy plans. This versatility ensures that different types of patients—from those requiring short-term intensive care to long-term home support—can benefit from its technology.

Additionally, the ventilator features a compact design, making it lightweight and portable. This ensures that it can be easily transported for both hospital use and home care without compromising on performance. The quiet operation of the ResMed VS III further enhances its usability, providing a more pleasant experience for both patients and caregivers.

In conclusion, the ResMed VS III stands out in the field of respiratory support devices due to its advanced technology, adaptability, and focus on patient comfort. By integrating connectivity features and a user-friendly interface, it empowers clinicians to deliver effective and personalized care.