Honeywell Manning EC-P2 installation manual APeriodic Maintenance, BSensor Life, CCalibration

Page 9
3 Maintenance

3 Maintenance

APeriodic Maintenance

It is essential that the test and calibration schedule be adhered to. Honeywell Analytics recommends the following maintenance schedule:

Response test prior to each use. Expose sensor to a sample of the target gas to verify proper sensor response and alarm functions.

Calibration should be performed with certified calibration gas every six months, and all tests must be logged.

Check filter material every six months. Replace if dirty or moist.

If the optional sensor keeper is used for extra SMART-CELLs, change the sensor keeper battery once every 6 months. Use only alkaline “C” size batteries.

BSensor Life

The electrochemical SMART CELL is extremely reliable, but the following can cause the cell chemicals to become depleted:

A period of time

Exposure to high temperatures

Exposure to high concentrations of the target gas

Exposure to high moisture

When the cell becomes depleted, the unit will give no indication of failure other than that the sensor will not respond. For this reason it is absolutely essential that these units be exercised with a gas sample on a regular basis.

Typical sensor life under normal use is two years or more. When the SMART-CELL will no longer calibrate properly, simply purchase a new SMART-CELL and replace.

CCalibration

Calibration is recommended every six months in normal use. Each SMART-CELL stores its calibration and alarm setpoint information on its own memory chip. When the SMART-CELL is plugged into the Manning EC-P2, this information is used by the sensor. The SMART-CELL can be field calibrated by the user or returned to Honeywell Analytics for factory calibration.

The calibration procedure requires the use of a bypass tee to allow the gas to be drawn into the flow cell without creating variable pressure or dilution problems. Calibration gas cannot be fed into the flow cell under pressure.

When applying calibration gas use the bypass tee as shown in Figure 4 on page 11. The flowrate of calibration gas should be high enough that at least

0.5SCFH is flowing out of the calibration tee. This can be measured with the supplied flowmeter and insures that air is not being drawn into the unit and diluting the calibration gas.

The unit may be zeroed without spanning it; however, never span the unit without first zeroing it.

The following procedure will calibrate the SMART- CELL.

To Enter ZERO / SPAN Mode

1)Push power button to turn unit on. Wait for power up sequence to finish (30 seconds).

2)Push and hold the unnamed button (second from left) first, then Push and hold the Menu button. Hold both buttons until the screen goes blank (2 seconds). Release both buttons and the unit is now in Programming Mode.

3)Push the SENS button and the unit is now in the zero/span mode.

Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved.

9

Image 9
Contents Manning EC-P2 Instruction and Installation ManualRelease G Draft Gas SensorNotices and Trademarks Copyright 2009 by Honeywell International IncRelease G July SymbolDefinition Symbol DefinitionsAbout This Document World Wide WebSerial number ContentsIntroduction SectionFigure 1 Basic Parts of the Manning EC-P2Sensor 1 System Description1 System Description continued System SpecificationsRepeatability ± 1% full scale Operating Ambient Temperature Range 0o to2 Operation AOperating the Sensorbutton must be held down for approximately three GAlarm Functions EResponse/Recovery TimeHDisplay Resolution 2 Operation continuedAPeriodic Maintenance 3 MaintenanceBSensor Life CCalibrationDTroubleshooting 3 Maintenance continuedUnit does not respond when keys are pressed Unit is in alarm with gas reading of 0 ppm CheckAAlarm Setpoints, Functions 4 Programming and Advanced FeaturesTo Enter CAUT / WARN / ALRM Mode Warning SetpointBDisplay Variables 4 Programming and Advanced Features continuedCSample Mode To Enter AVG / BLANK ModeDData Logging EC-P2software installationProgramming sample mode variables To Enter SAMP / MEAS / CLEAR ModeEAnalog Output Adjusting the RANGETo Stop the Data Logging Session Working with the “EC-P2”programAPPENDIX A SMART-CELLsAvailable 5 AppendicesRANGE PART NUMBERI N T E R F E R E N C E G A S E S 5 Appendices continued1.Limited Warranty 6 Limited Warranty2.Exclusions 4.Disclaimer of Unstated Warranties
Related manuals
Manual 28 pages 24.6 Kb

Manning EC-P2, EC-P2 specifications

The Honeywell EC-P2 is a sophisticated and versatile product designed for enabling efficient monitoring and management of various building systems. This advanced device is particularly popular among facility managers and operational teams for its exceptional features and robust technology that enhance the overall functionality of building automation.

One of the standout features of the Honeywell EC-P2 is its dual communication capability. It integrates seamlessly with both BACnet and Modbus protocols, enabling effective interoperability within diverse building management systems. This flexibility allows users to connect various devices and sensors, creating a unified and streamlined infrastructure.

The EC-P2 is equipped with advanced sensor technology that provides real-time data on numerous environmental parameters. By accurately monitoring variables such as temperature, humidity, and air quality, the device plays a crucial role in optimizing indoor comfort and energy efficiency. This real-time monitoring helps facility managers make informed decisions and deploy strategies that reduce energy consumption and operating costs.

Another notable characteristic of the Honeywell EC-P2 is its user-friendly interface. The device comes with an intuitive display that provides easy access to key parameters and system statuses. Users can easily configure settings and troubleshoot any issues directly through the interface, minimizing downtime and improving operational efficiency.

In terms of expandability, the EC-P2 offers multiple I/O options, making it versatile enough to handle various applications. This expandability allows for integration with additional sensors and control devices, ensuring that the system can grow alongside changing operational needs.

Notably, the EC-P2 is designed with security in mind. It incorporates robust security measures to safeguard data and ensure safe communication over the network. This focus on security is increasingly important in today’s connected environments, where cybersecurity threats pose significant risks.

Finally, the Honeywell EC-P2 stands out for its reliability and durability. Built with quality components and rigorous testing standards, it is engineered to function effectively in demanding environments. Whether used in commercial buildings, industrial settings, or educational institutions, the EC-P2 is a trusted solution for modern building management needs.

With its advanced features, wide-ranging compatibility, and focus on efficiency and security, the Honeywell EC-P2 is a pivotal component in creating intelligent, responsive, and sustainable building environments. Its integration capabilities and user-centric design make it an invaluable tool for optimizing operational performance in today's fast-evolving technological landscape.