ENDOSCOPE DISINFECTION GUIDE

VISERAPEF-V Transnasal Esophagovideoscope

Refer to companion ENDOSCOPE CLEANING GUIDE for cleaning information

HIGH-LEVEL DISINFECTION

Injection Tube

Channel Plug

Fill basin with disinfectant. Attach channel plug

 

 

 

While the endoscope is immersed,

and injection tube to previously cleaned

Use syringe to completely fill air, water and

disconnect cleaning accessories and use

endoscope. Immerse completely in disinfectant

suction channels with disinfectant solution.

a clean lint-free cloth to wipe away any

bubbles clinging to external surfaces.

solution.

 

 

 

 

 

 

 

RINSING

4

5

6

CAUTION: Follow disinfectant

 

 

 

manufacturer's rinsing instructions.

Soak all equipment according to the

Connect channel plug and injection tube.

Disconnect channel plug and injection

disinfectant manufacturer's recommended

Inject air to remove disinfectant from air,

tube. Immerse endoscope and

time and temperature. Use a timer to

water and suction channels. Remove

detached parts in water. Thoroughly

measure exact contact time.

endoscope from disinfectant.

rinse all external surfaces.

DRYING

7

8

9

Alcohol

Connect channel plug and injection tube to

Remove endoscope from rinse water. Use

Use syringe to flush 70% alcohol through

endoscope. Use a syringe to flush water

syringe to inject air through air, water and

air, water & suction channels. Inject air

through air, water and suction channels.

suction channels, expelling the rinse water.

to remove retained alcohol.

TRANSPORTATION

10

Disconnect all accessories from the endoscope.

 

Dry endoscope and accessories with an alcohol-

Hold all parts of the endoscope when

moistened cloth. Make sure equipment is

carrying. Do not let the ends of the

completely dry prior to storage.

endoscope swing freely.

IMPORTANT

Use a compatible disinfectant. Test for proper concentration per manufacturer's instructions.

Autoclave all autoclavable accessories.

Store endoscope hanging vertically with valves and caps removed, and with angulation lock in a “free” position.

Clean endoscope meticulously prior to disinfection.