 BRONCHOSCOPE DISINFECTION GUIDE

for OES® BF-40 Series BronchofiberscopesHIGH-LEVEL DISINFECTIONRINSINGDRYINGTRANSPORTATIONNOMENCLATURE BRONCHOSCOPE DISINFECTION GUIDE

for OES® BF-40 Series Bronchofiberscopes

Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information

WARNING: This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.

HIGH-LEVEL DISINFECTION

1

2

3

4

Suction Cleaning Adapter

Fill basin with disinfectant. Attach a suction cleaning adapter (MAJ-222) to the previously cleaned bronchoscope. Immerse completely in disinfectant solution.

Attach a syringe to the adapter. Pull the plunger of the syringe to fill the instrument/suction channels and channel cleaning adapter with disinfectant solution.

While immersed, detach the adapter from the bronchoscope. Use a clean, lint-free cloth to wipe away any bub- bles clinging to external surfaces.

Soak all equipment according to the time and temperature recommended by the disinfectant manufacturer. Use a timer to mea- sure the exact contact time.

RINSING

5

 

6

 

 

7

To Suction Pump

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Connect the suction cleaning adapter

Disconnect the suction cleaning

Connect suction cleaning adapter

and syringe. Inject air to remove

adapter and immerse the broncho-

to bronchoscope and to a tube

disinfectant from the instrument/suction

scope and adapter in sterile water.

from a suction pump. Aspirate

channels, then remove the broncho-

Using a sterile, lint-free cloth, thor-

water through channels for 30

scope from the disinfectant.

oughly rinse and wipe all surfaces.

seconds. Remove from water and

 

 

aspirate air through channels for

 

 

60 seconds.

8

To Suction Pump

Alcohol

Immerse the bronchoscope's distal end in 70% alcohol and aspirate alcohol for 5 seconds. Remove the distal end from the alcohol and aspirate air for 20 seconds.

DRYING
TRANSPORTATION

9

10

Turn the suction pump off. Hold the control

Dry the bronchoscope and adapter with an

 

section with the openings pointing down and

alcohol-moistened cloth. Make sure that the

Hold all parts of the endoscope when carrying. Do

disconnect the suction cleaning adapter.

equipment is completely dry prior to storage.

not let the ends of the bronchoscope swing freely.

 

NOMENCLATURE

 

 

Single-Use Suction Valve (MAJ-209)

Universal Cord

 

Light Guide

 

 

 

Suction Cylinder

 

 

 

 

 

Single-Use Biopsy Valve (MAJ-210)

 

 

 

Instrument/Suction Channel Port

Venting Connector

Control Section

Boot

Insertion Tube

 

 

UP/DOWN Angulation Control Lever

 

Bending Section

Distal Tip

 

 

IMPORTANT

Wear appropriate personal protective equipment during reprocessing.

Autoclave all autoclavable accessories.

Store the bronchoscope hanging vertically with all valves and caps removed.

Use a compatible disinfectant and test it for proper concentration as recommended by the manufacturer.

Do not use the carrying case to store "patient-ready" endoscopes. Any bronchoscope placed in the carrying case must be reprocessed prior to patient use.

Clean the bronchoscope meticulously prior to disinfection.

Clean the bronchoscope meticulously prior to disinfection.© 2004 Olympus America Inc. / Printed in USA 01/04 — PWC-100020