1.0 Description
The 3M™ Benchtop Air Ionizer 963E is a
2.0 Performance
The Ionizer 963E reduces a static charge of ± 1000V to ± 100V in less than two seconds (the discharge time) at a distance of one foot (30 cm) using the high fan speed. Testing is performed in accordance with the ionization standard ANSI/ ESD S3.1.
3.0 Power Requirements
The Ionizer 963E requires 24VDC power, which is supplied through a Mini DIN connector on the back of the unit. The Ionizer 963E is packaged with a switching power supply, capable of converting
4.0Installation
4.1The Ionizer 963E mounts easily in a variety of positions using the
provided tilting bracket/stand. Place the unit on the work surface and point it at the area or object to be neutralized. Alternatively, the mounting bracket may be attached directly to or above the workstation, or on another supporting structure. Please note that placement of the Ionizer 963E is important in determining its effectiveness. The distance from the target object, and fan speed affect the ionizer’s performance. As distance increases or fan speed is reduced, the discharge time will increase.
4.2Connect the switching power supply to the Ionizer 963E using the Mini DIN connector. Then, using the appropriate electrical power cord, connect the switching power supply to an electrical outlet.
4