3M M5560 Troubleshooting, Curing Performance Test, Repair, Specifications, Condition, Remedy

Models: M5560

1 8
Download 8 pages 58.37 Kb
Page 7
Image 7

Troubleshooting:

Condition

Main indicator light inoperative

Curing bulb inoperative

No time signal tone

Reduced curing performance

Handpiece turns on, immediately hear 5 beeps and handpiece shuts off

Unit overheats. Five consecutive beeps, bulb shuts off during curing cycle

Remedy

1.Check electrical supply.

2.Send unit to an authorized service center.

1.Replace bulb.

2.Send unit to an authorized service center.

1.Send unit to an authorized service center.

1.Check intensity with hand held radiometer.

2.Clean light guide exit tip. Consult cleaning instructions.

3.Clean optical filter in nose cone with a soft cotton swab.

4.Check intensity output.

5.Check optical filter in nose cone for chips or cracks. If chips or cracks are present on optical filter, send unit to an authorized service center.

6.Replace light guide.

7.Replace bulb.

8.Check intensity output.

9.Send unit to an authorized service center.

1.Fan possibly blocked, therefore send unit to an authorized service center.

1.Let unit cool down for 10 minutes.

Restart bulb. If, after a few minutes the bulb shuts off again after 5-beep signal, or if the bulb will not restart, send the unit to an authorized service center.

Curing Performance Test

Lay a clean polyester matrix band on white paper and place a clear or white plastic form on the band. The form should be a cylindrical hole 3-6 mm diameter through material 2.0-2.5 mm thick. Pack restorative material into the form, cover with polyester and press flat. Direct the curing light at the top surface (from less than 2 mm) of the sample and cure for the recommended time. Remove the polyester strips and examine the condition of cure. The bottom surface should appear glassy and resist scratching by a sharp explorer, indicating adequate cure. (Consult manufacturer’s instructions for other criteria for correct cure).

If the surface is soft or dull, repeat the test procedure using incrementally longer curing times until adequate cure is attained, and adjust curing time accordingly. If the adjusted time exceeds 200% of normal, verify that the restorative material has not degenerated, or correct curing light problems as directed in the Troubleshooting table.

Repair

Obtain qualified repair service only from 3M ESPE authorized service facilities. Contact:

3M Health Care Service Center Suite 200 Bldg. 502

3350 Granada Ave. North Oakdale, MN 55128 (800) 292-6298

Replacement Parts and Accessories - Can be purchased from authorized dealer

Item

3M ESPE Stock Number

Bulb/12V, 75W

78-8131-1662-7

Light guide - 8 mm diameter

78-8131-1657-7

Light guide - 13 mm diameter

78-8131-1658-5

Mounting Kit

78-8131-1663-5

Eyeshield - 8 mm

78-8131-1660-1

Eyeshield - 13mm

78-8131-1661-9

Specifications

Note: This manual is used for all versions of the 3M ESPE Elipar 2500, M5560.

Operation of equipment - Continuous Duty

-Unit will operate continuously for 5 minutes before automatic shutoff, at which time the unit may be restarted immediately