Troubleshooting:
Condition
Main indicator light inoperative
Curing bulb inoperative
No time signal tone
Reduced curing performance
Handpiece turns on, immediately hear 5 beeps and handpiece shuts off
Unit overheats. Five consecutive beeps, bulb shuts off during curing cycle
Remedy
1.Check electrical supply.
2.Send unit to an authorized service center.
1.Replace bulb.
2.Send unit to an authorized service center.
1.Send unit to an authorized service center.
1.Check intensity with hand held radiometer.
2.Clean light guide exit tip. Consult cleaning instructions.
3.Clean optical filter in nose cone with a soft cotton swab.
4.Check intensity output.
5.Check optical filter in nose cone for chips or cracks. If chips or cracks are present on optical filter, send unit to an authorized service center.
6.Replace light guide.
7.Replace bulb.
8.Check intensity output.
9.Send unit to an authorized service center.
1.Fan possibly blocked, therefore send unit to an authorized service center.
1.Let unit cool down for 10 minutes.
Restart bulb. If, after a few minutes the bulb shuts off again after
Curing Performance Test
Lay a clean polyester matrix band on white paper and place a clear or white plastic form on the band. The form should be a cylindrical hole
If the surface is soft or dull, repeat the test procedure using incrementally longer curing times until adequate cure is attained, and adjust curing time accordingly. If the adjusted time exceeds 200% of normal, verify that the restorative material has not degenerated, or correct curing light problems as directed in the Troubleshooting table.
Repair
Obtain qualified repair service only from 3M ESPE authorized service facilities. Contact:
3M Health Care Service Center Suite 200 Bldg. 502
3350 Granada Ave. North Oakdale, MN 55128 (800)
Replacement Parts and Accessories - Can be purchased from authorized dealer
Item | 3M ESPE Stock Number |
Bulb/12V, 75W | |
Light guide - 8 mm diameter | |
Light guide - 13 mm diameter | |
Mounting Kit | |
Eyeshield - 8 mm | |
Eyeshield - 13mm |
Specifications
Note: This manual is used for all versions of the 3M ESPE Elipar 2500, M5560.
Operation of equipment - Continuous Duty
-Unit will operate continuously for 5 minutes before automatic shutoff, at which time the unit may be restarted immediately
7