60
The following commands can not be used during data storage.
Q Query command for weighing data.
S Query command for stable weighing data.
SI Query command for weighing data.
SIR Query command for continuous weighing data.
Parameter settings for each output mode are as follows:
Item
Mode
Data output
mode Auto print
polarity,
difference
Data memory
function Interval
time
Key mode prt 0 Not used data 2
Auto print mode A prt 1 data 2
Auto print mode B prt 2
ap-a 0-2
ap-b 0-2 data 2
Not used
Interval memory mode prt 3 Not used data 2 int 0-8
Set each item, depending on the situation, as follows:
No d-no 0 No 5-td 0
Data number Yes d-no 1 Time only 5-td 1
No 5-id 0 Date only 5-td 2 ID number Yes 5-id 1
Time and date
Both 5-td 3
With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets.
1 Press and hold the SAMPLE key until ba5fnc is displayed, then release the key.
2 Press the SAMPLE key several times to display dout .
3 Press the PRINT key.
4 Press the SAMPLE key three times to display data 0 .
5 Press the RE-ZERO key to display data 2 .
6 Press the PRINT key to store the setting.
7 Press the CAL key to return to the weighing mode.