Table
SPECIFICATIONS
DC Input Rating:
Current: 0 to 30 A
Voltage: 3 to 60 V (minimum dc operation from 0 to 2 V for 0 to 30 A)
Power: 150 W at 40 °C (derated to 112 W at 55 °C)
A. OPERATING CHARACTERISTICS | B. DERATED CURRENT DETAIL | |
Constant Current Mode: |
|
|
Ranges: | 0 to 3 A; and 0 to 30 A | |
Accuracy: | (after 30 second wait): ± 0.1% ± 40 mA (both ranges) | |
Resolution: | 0.8 mA (3 A range); 8 mA (30 A range) | |
Regulation: | 10 mA (both ranges) | |
Temperature Coefficient: | 100 ppm/°C ± 3 mA/°C (both ranges) | |
Constant Resistance Mode: |
|
|
Ranges: | 0.067 to 2 Ω; 2 Ω to 2 kΩ; and 20 Ω to 10 kΩ | |
Accuracy: | ± 0.8% ± 16 m Ω with ≥ 6 A at input (2 Ω range); | |
| ± 0.3% ± 5 mS with ≥ 6 V at input (2 k and 10 kΩ ranges) | |
Resolution: | 0.54 mΩ (2 Ω range); 0.14 mS (2 kΩ range); 0.014 mS (10 kΩ range) | |
Regulation: | 10 mV with remote sensing (2 Ω range); 10 mA (2 k and 10 kΩ ranges) | |
Temperature Coefficient: | 800 ppm/°C ± 0.8 mΩ/°C (2 Ω range); | |
| 300ppm/°C ± 0.5 mS/°C (2 k and 10 kΩ ranges) | |
Constant Voltage Mode: |
|
|
Range: | 0 to 60 V |
|
Accuracy: | ± 0.1% ± 50 mV | |
Resolution: | 16 mV |
|
Regulation: | 5 mV (remote sense); 40 mV (local sense) | |
Temperature Coefficient: | 100 ppm/°C ± 5 mV/°C |
2