5-16
Servicing
Troubleshooting
3If the green LED is on, the current source driving the pump laser is operating
correctly. If the yellow LED is on, the pump laser’s temperature stabilization
loop is not functioning; the current to the pump laser is automatically turned
off which turns the green LED off.
4If the yellow LED is off, check ribbon cable W12.