58 E1442A Command Reference   Chapter 3
DISPlay
The DISPlay subsystem monitors the channel state of a selected modu le 
(or card) in a switchbox. The DISPlay command subsystem only 
operates with an RS-232 terminal connected to the E1406 Command 
Module’s RS-232 port. These commands control the display on the 
terminal, and would in most cases be typed directly from the terminal  
keyboard. 
However, it is possible to send these commands over the GPIB interface 
and control the terminal’s display. In this case, care must be taken that 
the instrument receiving the DISPlay command is the same one that is 
currently selected on the terminal. Otherwise, the GPIB command will 
have no visible effect.
Subsystem Syntax DISPlay
:MONitor
:CARD <number>|AUTO
:CARD?
[:STATe]<mode>
[:STATe]?
DISPlay:MONitor:CARD
DISPlay:MONitor:CARD <number>|AUTO selects the module in a 
switchbox to be monitored. You must use DISP:MON:STAT ON to 
actually display the monitored module state to the RS-232 terminal.
Parameters
Comments Selecting a Specific Module to be Monitored: Send the card number in a 
switchbox with the DISPlay:MONitor:CARD command.
Selecting the Present Module to be Monitored (AUTO): Use the 
DISPlay:MONitor AUTO command to select the last module addressed 
by a switching command (e.g., [ROUTe:]CLOSe).
*RST Condition: DISPlay:MONitor:CARD AUTO
Example Select Module #2 in a Switchbox for Monitoring
',63021&$5' Selects module #2 in a 
switchbox
Name Type Range of Values Default Value
<number | AUTO> numeric 1 - 99  AUTO