Aiphone NH-1SA-AC installation instructions Nurse Center System, Wiring Diagram, Patient Stations

Models: NH-2SA-AC

1 2
Download 2 pages 15.75 Kb
Page 1
Image 1
Nurse Center System

NHX

Nurse Center System

NH-1SA-AC NH-2SA-AC

Patient Stations

Single and Dual Patient Stations for All Call, and when sharing a common Corridor Lamp

- INSTRUCTIONS -

Some applications for Nurse Call systems require All Call. By including the NH-50AC adaptor and the "AC" model patient stations, a station button on the master can be used to call all of the patient stations simultaneously. An additional wire is required to each patient station. In addition, if there is more than one patient station sharing a corridor lamp, this model patient station will allow that configuration to be used.

NOTE: An NHX system installed with an NH-1SA-AC or NH-2SA-AC included does not carry the UL 1069 listing.

WIRING DIAGRAM:

Two Patient Stations in one room sharing a bathroom and corridor lamp

NHX-80X

11

12

Trunkage Line 13

Connection 14

Terminals

15

16

21

22

23

24

25

26

~

51

52

53

54

55

56

 

PS-2420UL

A+

+

A-

-

NH-1SA-ACNH-2SA-AC

CN1:

CN1:

Brn

Brn

Red

Red

Org

Org

Yel

Yel

Grn

Grn

Blu

Blu

Blk

Blk

CN2:

CN2:

Brn

Brn

Red

Red

Org

Org

Yel

Yel

Grn

Grn

Blu

Blu

Vio

Vio

 

* Bath call will annunciate on

 

one position at the master

 

station, according to which

 

patient station it is wired to.

APPLICATION #1:

All Call feature added, where an announcement can be made from the master station to all patient stations simultaneously.

APPLICATION #2:

2 patient stations in a room with a common bathroom and one corridor lamp outside the room.

To NH-50AC

NHR-4A

 

Brn

Corridor Lamp

Red

(Shared by both patient stations)

 

Org

 

NHR-7A

 

 

Yel

Bathroom Pullcord

Grn

(Wired to one station only)

 

 

Blu

B+

+

B-

-

 

PS-2420UL

NOTES:

1.All Call requires one of the station buttons to be used for All Call. Therefore, the standard system can have up to 49 stations, and the expanded system can have up to 79 stations.

2.Only wiring concerning the NH-1SA-AC and NH-2SA-AC is shown here. For complete installation and wiring information, please see the NHX Installation instructions.

Aiphone Communication Systems

TOLL FREE TECHNICAL SUPPORT:

 

1700 130th Ave. N.E.

(800) 692-0200

 

Bellevue, WA 98005

TOLL FREE FAX LINE:

 

(425) 455-0510

(800) 832-3765

NH-1/2SA-AC Instr.

FAX (425) 455-0071

E-MAIL: tech-serv@aiphone.com

0103js

Page 1
Image 1
Aiphone installation instructions Nurse Center System, Wiring Diagram, NH-1SA-AC NH-2SA-AC, Patient Stations