Packing List
Examine the shipping carton and the contents for physical damage. The following items should be in the shipping carton:
1.FBDA module
2.This instruction manual
If any of these items are damaged or missing, contact B&B Electronics immediately.
Software
Software is not included with the FBDA module. This device is used in conjunction with a data acquisition device. All of B&B Electronics’ data acquisition modules come complete with software that you can modify to meet your exact needs.
Calibration
The FBDA is calibrated at the factory. The excitation voltage is set to 5VDC ± 0.05VDC and the gain is calibrated at 80 ±
2.The process of calibrating the FBDA and your sensor will vary depending on the type of sensor being used. An example calibration procedure is covered in Chapter 3.
FBDA Specifications |
|
Size: | 4.8 x 2.2 x 0.9 in |
Maximum Input Voltage from Sensor: | 440mV (using Gain = 25) |
Output Voltage Range: | 0.003V to 11VDC |
Gain: | 25, 40, 80, |
Maximum Gain Error: | 5% of gain setting max. |
Input Offset Voltage: | 150∝ V max. |
Input Offset Voltage Drift: | 1.5∝ V/oC |
Excitation Voltage: | 0.5VDC to +11VDC |
Operating Temperature Range: | 0 to 70oC |
Power Requirements: | |
| not include current draw |
| from excitation voltage.) |
2Documentation Number FBDA0797 Manual
B&B Electronics
PH (815)