CLOCK – Sets the monitor’s date and time. Use the Keypad to enter the correct date and time, and then press ENTER to save those values and return to the previous screen. Note that time is displayed in a
24 hour format, while the date is displayed as mm/dd/yy.
SILENCE – Used to enter a length of time for which the internal audible alarm and the external alarm are turned OFF when the front panel SILENCE button is pressed. The factory default is 300 seconds (5 minutes). If the cause of the alarm/fault has not been cleared at the end of this time period, the internal audible alarm and the external alarm device are reactivated. Use the Keypad to enter the desired time period, and then press ENTER to save that value and return to the previous screen.
SQUELCH – Sets a value of between 0.0 and 99.9 ppm that prevents the display of measurements below that value. Factory default is 0 ppm. For example, if the squelch setting is set to 50 ppm, then the monitor will not display measurements that are below that value. Use the Keypad to enter the desired value, and then press ENTER to save that value and return to the previous screen.
ZONETEMP – Used to enter the temperature of the area being monitored in °C, thus giving a more accurate ppm reading. The factory default is 25°C. Use the Keypad to enter the desired temperature, and then press ENTER to save that value and return to the previous screen.
CAL – Used to change the monitor’s calibration factor. This function is to be used ONLY with instructions from a Bacharach Service Center. Factory default is 1.000.
SET DATE & TIME 11/10/03 15:30
SILENCE TIMEOUT 0300 sec
SQUELCH BELOW 50.0ppm
TEMP AT ZONE = 25 degC
R717 CALFACTOR
1.000
14.72* AMB14.81 0.09dif <0000>
Instruction | 17 |