LCD Lift Codes
Listed below are codes that the lift controller outputs during lift operation. The codes will be displayed on an LCD screen located on the lift control board inside the pump module. See the Manual Operating Instruc- tions in the operator's manual for pump cover removal instructions.
Non-Flashing Numbers
01 – Platform stowed
02 – Platform unfolding
03 – Platform unfolding paused ²3ODWIRUPDWÁRRUOHYHO
05 – Platform beginning to lower
06 – Platform lowering (threshold cannot be occupied from this point down)
07 – Outer barrier moving to horizontal position
08 – Platform at ground level
09 – Outer barrier moving to vertical position
10 – Platform raising
²3ODWIRUPUDLVLQJSDXVHGDWÁRRU
12 – Platform folding (limited pressure)
13 – Platform folding (full pressure)
14 – Timed fold (cinching lift tite) or
15 – Platform folding stopped
16 – Paused fold
17 – Platform between ground and 3” above ground
18 – Platform above 3”
19 – Outer barrier moving to horizontal postion ²,OOHJDOIXQFWLRQQRWGHÀQHG
29 – Interlock fault not recognized (or has been cleared but a motion button is still pressed)
30 – Platform location unknown
31 – Platform location transition state; attempting to locate position
35 – Two or more motion buttons are being pressed
36 – The retention belt cannot be buckled while trying to fold or unfold
37 – Motion button being pressed is not a valid motion 50 – Outer barrier is not up above inboard barrier
locked position
51 – Threshold is occupied when platform is 1” or more
EHORZÁRRUOHYHO
52 – Inner rollstop is not up and locked below inner rollstop locked position
53 – Inner rollstop occupied sensor is not activated
EHWZHHQÁRRUDQGLQQHUUROOVWRSXSSRVLWLRQ 54 – Outer barrier is occupied before it is up
55 – Outer barrier is not latched when above the inner rollstop locked position (Millennium only)
56 – Outer barrier is not up and latched and bridge switch did not deactivate
57 – Outer barrier is not up and latched and ground detect switch did not deactivate (Century and Vista only)
58 – Outer barrier is not up and latched and the platform is 3” above the ground
59 – Outer barrier is not up after pausing platform travel 60 – The kickout gas springs are worn, replace before
using
75 – Low voltage detected; must turn off power switch to reset LCD
77 – Vehicle secure interlock has not been activated
90 – Position will be set if you keep holding the button until it beeps
91 – Position is out of a predetermined acceptable range
RIÁRRUSRVLWLRQ
92 – Bridge switch is not made, needs adjusting
93 – Inner rollstop occupied switch is not made, position needs to be moved or switch should be adjusted
²2XWHUEDUULHULVQRWPDGHÀ[DQGWU\DJDLQ
95 – Outer barrier latch is not made (check for jumper on Century and Vista lifts, check latch on Millennium lifts)
99 – Controller program is not valid; replace controller
Flashing Numbers
Flashing 65 – Unfold button is pressed Flashing 66 – Fold button is pressed Flashing 67 – Down button is pressed Flashing 68 – Up button is pressed Flashing 69 – Bridge switch is activated Flashing 70 – Outer barrier latch switch is activated Flashing 71 – Ground detect switch is activated Flashing 72 – Outer barrier up switch is activated Flashing 73 – Inner rollstop up switch is activated Flashing 74 – Inner rollstop occupied switch is activated Flashing 76 – Outboard barrier occupied switch is
activated
Flashing 78 – Threshold tape switch “A” is activated Flashing 79 – Threshold tape switch “B” is activated Flashing 80 – Position set button is pressed
Page 7