7. Specifications
7.1 Main Specifications
Sensor Unit
Purpose | General radiography |
Pixel size | 160 × 160 µm |
Image matrix size | 2208 × 2688 pixels |
Number of pixels | Approx. 5.9 million pixels |
Dynamic range | Approx. 80 dB |
Resolution | 3.1 lp/mm |
Gray scale | |
Environmental requirements | Operation: |
| Temperature: +5 to +35 °C |
| Humidity: 30 to 75 %RH (no condensation) |
| Storage and transportation: |
| Temperature: |
| Humidity: 10 to 60 %RH (no condensation) |
| Atmospheric pressure: 700 to 1060 hPa |
Power | Supplied from the power box. |
Dimensions and mass | 480 (W) × 481 (H) × 15 (D) mm, |
| 3.4 kg (excluding the sensor cable) |