BP-1MY-6
To stop an ongoing measurement operation, press the | button. This | |
causes the Base Unit to deflate the cuff and then go into standby. | ||
Pressing the | button also causes the Base Unit to deflate the cuff | |
and turn off. |
|
|
You can make copies of the “Blood Pressure Record” at the back of this manual and use them to keep track of changes in your blood pressure and pulse measurements.
ERROR MESSAGES
When any of the errors described below occurs, the Base Unit deflates the cuff and then waits for further button operation or data communication with the Watch.
Message |
| Cause | Action | |
Err 01 | • The Base Unit cannot | • Do not move or speak | ||
|
| detect a suitable pulse. | while measurement is in | |
|
|
| progress. | |
Err 02 | • | setting is too low. | ||
• Make sure the air tube is | ||||
| • Noise due to body | |||
| not under your arm. | |||
|
| movement. |
| |
Err 03 | • The cuff is not connected | • Check the air tube connec- | ||
|
| to the Base Unit properly, | tion to the Base Unit. | |
|
| or a hole in the cuff. | • Tighten the cuff so there is | |
| • The cuff is too loose on the | no space between your | ||
|
| arm. | arm and the cuff. | |
|
|
| • If there is air leaking from | |
|
|
| the cuff because it is old or | |
|
|
| it has a hole in it, replace | |
|
|
| the cuff with a new one. | |
|
|
|
| |
Err 04 | • | value is too low. | • Do not move or speak | |
| • Noise due to body | while measurement is in | ||
|
| movement. | progress. | |
| • The cuff is too loose on the | • Make sure the air tube is | ||
|
| arm. | not under your arm. | |
|
|
| • Tighten the cuff so there is | |
|
|
| no space between your | |
|
|
| arm and the cuff. | |
|
|
| ||
Err 44.0 | • Increase in cuff pressure | • Take care so no pressure | ||
(kPa) |
| due to body movement. | is applied to the cuff while | |
Err 330 |
|
| measurement is in | |
(mmHg) |
|
| progress. | |
|
|
| ||
|
|
|
|
The appearance of the low battery indicator shown below causes the Base Unit to cancel any ongoing measurement or communication operation and wait for further button operation.
Indicator | Cause | Action |
| • Battery power is low. | Replace the Base Unit |
|
| batteries. |
|
|
|
TROUBLESHOOTING
Check the points noted below whenever you experience operational problems with the Base Unit.
Problem | What to Check and What to Do |
Nothing happens when | • Is battery power low? |
you press the | → Replace the batteries with a full set of |
button. | new ones. |
|
|
| • Are batteries loaded correctly? |
| → Check to make sure that batteries are |
| loaded correctly. |
|
|
Inflation is performed, but | • Is the cuff’s plug connected correctly to the |
the cuff does not inflate. | Base Unit? |
| → Check the plug for proper connection. |
|
|
| • Is there a hole in the cuff? |
| → Contact the store or dealer where you |
| purchased the Base Unit about replacing |
| the cuff. |
|
|
If the actions recommended above do not correct your problem, the Base Unit itself may be malfunctioning. In this case, contact the store or dealer where you purchased the Base Unit.
PRECAUTIONS DURING USE
•Do not leave the Base Unit in areas exposed to direct sunlight or near a heater.
•Do not expose the Base Unit to severe vibration or impact.
•To clean the exterior of the Base Unit, wipe it with a dry cloth or with a cloth moistened with a weak solution of water and a mild neutral detergent. Wring all excess liquid from the cloth before wiping. Never use thinner, benzene, or other volatile agents for cleaning. Doing so can discolor or damage the case.
•Do not submerse the cuff in water or launder it.
•Be sure to remove the batteries from the Base Unit if you do not plan to use it for a long time.
CASIO COMPUTER CO., LTD. assumes no responsibility for any loss, or any claims by third parties that may arise through the use of the Base Unit.