Stopping a Measurement

To stop an ongoing measurement operation, press the

button. This

causes the Base Unit to deflate the cuff and then go into standby.

Pressing the

button also causes the Base Unit to deflate the cuff

and turn off.

 

 

Recording Blood Pressure Readings

You can make copies of the “Blood Pressure Record” at the back of this manual and use them to keep track of changes in your blood pressure and pulse measurements.

ERROR MESSAGES

When any of the errors described below occurs, the Base Unit deflates the cuff and then waits for further button operation or data communication with the Watch.

Message

 

Cause

Action

Err 01

The Base Unit cannot

Do not move or speak

 

 

detect a suitable pulse.

while measurement is in

 

 

 

progress.

Err 02

setting is too low.

Make sure the air tube is

 

Noise due to body

 

not under your arm.

 

 

movement.

 

Err 03

The cuff is not connected

Check the air tube connec-

 

 

to the Base Unit properly,

tion to the Base Unit.

 

 

or a hole in the cuff.

Tighten the cuff so there is

 

The cuff is too loose on the

no space between your

 

 

arm.

arm and the cuff.

 

 

 

If there is air leaking from

 

 

 

the cuff because it is old or

 

 

 

it has a hole in it, replace

 

 

 

the cuff with a new one.

 

 

 

 

Err 04

value is too low.

Do not move or speak

 

Noise due to body

while measurement is in

 

 

movement.

progress.

 

The cuff is too loose on the

Make sure the air tube is

 

 

arm.

not under your arm.

 

 

 

Tighten the cuff so there is

 

 

 

no space between your

 

 

 

arm and the cuff.

 

 

 

Err 44.0

Increase in cuff pressure

Take care so no pressure

(kPa)

 

due to body movement.

is applied to the cuff while

Err 330

 

 

measurement is in

(mmHg)

 

 

progress.

 

 

 

 

 

 

 

The appearance of the low battery indicator shown below causes the Base Unit to cancel any ongoing measurement or communication operation and wait for further button operation.

Indicator

Cause

Action

 

Battery power is low.

Replace the Base Unit

 

 

batteries.

 

 

 

TROUBLESHOOTING

Check the points noted below whenever you experience operational problems with the Base Unit.

Problem

What to Check and What to Do

Nothing happens when

Is battery power low?

you press the

Replace the batteries with a full set of

button.

new ones.

 

 

 

Are batteries loaded correctly?

 

Check to make sure that batteries are

 

loaded correctly.

 

 

Inflation is performed, but

Is the cuff’s plug connected correctly to the

the cuff does not inflate.

Base Unit?

 

Check the plug for proper connection.

 

 

 

Is there a hole in the cuff?

 

Contact the store or dealer where you

 

purchased the Base Unit about replacing

 

the cuff.

 

 

If the actions recommended above do not correct your problem, the Base Unit itself may be malfunctioning. In this case, contact the store or dealer where you purchased the Base Unit.

PRECAUTIONS DURING USE

Do not leave the Base Unit in areas exposed to direct sunlight or near a heater.

Do not expose the Base Unit to severe vibration or impact.

To clean the exterior of the Base Unit, wipe it with a dry cloth or with a cloth moistened with a weak solution of water and a mild neutral detergent. Wring all excess liquid from the cloth before wiping. Never use thinner, benzene, or other volatile agents for cleaning. Doing so can discolor or damage the case.

Do not submerse the cuff in water or launder it.

Be sure to remove the batteries from the Base Unit if you do not plan to use it for a long time.

CASIO COMPUTER CO., LTD. assumes no responsibility for any loss, or any claims by third parties that may arise through the use of the Base Unit.