DESCRIPTION OF DISPLAY MARKS

DISPLAY MARK

CONDITION / CAUSE

CORRECTIVE ACTION

 

The cuff was inflated to more than 280

Exhaust the air with in the cuff.

 

mmHg.

 

 

 

 

 

 

Pressurization is not possible

After winding the cuff over again, re-

 

Cuff winding has not been performed

 

peat measurement.

 

correctly.

 

 

 

The pulse could not be measured ac-

After again winding the cuff, conduct

 

curately.

the measurement without moving.

 

 

 

 

The arm moved during measurement.

Repeat the measurement without mov-

 

ing the arm.

 

 

 

 

 

 

Device has a trouble.

Ask purchased shop.

 

 

 

 

Appears when the pulse could not be

Repeat the measurement without mov-

 

measured.

ing the arm.

 

 

 

 

Appears when the batteries become

Replace the batteries.

 

weak.

 

 

Device has a trouble.

Ask purchased shop.