Cornelius Series 200, Series 525, Series 725 Maintenance, Sanitizing And Cleaning Procedure

Models: Series 725 Series 200 Series 525

1 22
Download 22 pages 53.91 Kb
Page 13
Image 13
MAINTENANCE

Self-Contained Flaked Ice Maker Service Manual

MAINTENANCE

IMPORTANT: THE FOLLOWING MAINTENANCE SHOULD BE PERFORMED AT LEAST EVERY SIX MONTHS ON FLAKED ICE MACHINES.

1.Check power supply with machine running for proper voltage.

2.Check water level in the float tank reservoir. Water level should be maintained at the top of the evaporator. Adjust if necessary. (See illustration and adjustment procedure)

3.Clean the air-cooled condenser coil with a stiff brush or vacuum cleaner (See procedure).

CAUTION: Condenser cooling fins are sharp, use care when cleaning.

4.Clean the ice storage bin and flush the bin drain at least once a month.

5.If a water conditioner is installed in the inlet water line, change, replace, or clean the filter, strainer or cartridge as required.

6.If heavy mineral deposits on the auger and evaporator shell are encountered due to bad local water conditions, follow sanitizing and cleaning procedure.

7.Loosen hold-down cam locks and remove gearmotor assembly.

8.Check thrust washer; replace if noticeably worn.

9.Lift out auger and examine for wear. The corkscrew auger guide bushing pressed into the drive block should be checked for wear. Replace if loose or if worn flat with auger drive block. If the Helix auger on the corkscrew auger round bar becomes flat on the inside more than 1/8 of an inch over a length of two inches or more it should be replaced.

NOTE: HELIX AUGERS DO HAVE MACHINED FLAT RELIEF SURFACES. DON’T CONFUSE THEM WITH WORN FLAT AREAS.

Check the insert in the bottom ring of the Helix auger and replace if excessively worn.

10.Check shell vertical strips for wear. Replace the shell if excessive wear is shown.

11.Check O-Ring, replace if worn or cut.

12.Re-assemble, steps 7 through 11.

CAUTION: In re-assembling the auger gearmotor, the hold down clamps must be tight and secure. In re-installing the evaporator shell, be absolutely sure that the “o” ring is not pinched off as this would cause a water leak around the base of the evaporator. Lubricate the “o” ring with food grade lubricant before re-assembling shell.

13.Check for alignment of ice chute. Make sure chute gasket is not blocking path of ice flow.

14.Check bin thermostat operation. In the mid-range position the bin thermostat will open at 42° and has a 6° differential.

SANITIZING AND CLEANING PROCEDURE

1.Turn unit off at switch in upper rear right side panel.

2.Turn water off and remove water hose from bottom of float chamber inside of ice bin and allow to drain from the evaporator via the hose end; or

3.Remove float chamber cover and while holding float up to prevent more water from entering the float chamber, remove water hose from float chamber and proceed to drain the float chamber and evaporator.

4.With water still off, restore water hose to float chamber and add 1/2 oz. of ”sanitizer” (see note below). Turn water on.

5.Remove the bin door, float chamber cover, ice chute trim cover, two thumb screws and chute trim gasket.

6.Using soap, hot water, and a non-metallic bristle brush or plastic scouring pad, scrub the parts removed in step 4 as well as the interior of the ice bin, ice chute, thermowell, door tracks, ice bin top, and around bin opening.

© 2004, IMI Cornelius Inc.

- 9 -

Publication Number: 166240010SER

Page 13
Image 13
Cornelius Series 200, Series 525, Series 725 service manual Maintenance, Sanitizing And Cleaning Procedure