Drive Medical Design 2850-18, 2850-20 manual Forward

Models: 2850-20

1 31
Download 31 pages 8.09 Kb
Page 4
Image 4
Manual background

FORWARD

Guidance and manufacturer’s declaration - electromagnetic emissions information

The Trident Image Intensifier is intended for use in the electromagnetic environment specified below. The customer or the user of the Trident Image Intensifier should assure that it is used in such an environment.

IMMUNITY test

IEC 60601

Compliance level

Electromagnetic environment - guidance

 

Test level

 

 

Electrostatic

± 6 kV contact

± 6 kV contact

Floors should be wood, concrete or ceramic

Discharge (ESD)

± 8 kV air

± 8 kV air

tile. If floors are contain synthetic

IEC 61000-4-2

 

 

material, the relative humidity should be at

 

 

 

least 30%.

Electrical fast

± 2kV for power

Not applicable

Mains power quality should be that of a

Transient/burst

Supply lines

 

typical commercial or hospital environment.

IEC 61000-4-4

± 1kV for

 

 

input/output lines

 

 

Surge

± 1 kV line(s) to

Not applicable

Mains power quality should be that of a

IEC 61000-4-5

line(s)

 

typical commercial or hospital environment.

 

± 2kV line(s) to

 

 

 

earth

 

 

Voltage dips,

<5% U T

 

Mains power quality should be that of a

short

(>95% dip in U T)

 

typical commercial or hospital environment.

interruptions and

for 0.5 cycle)

 

If the user of the (ME EQUIPMENT or ME

voltage variations

 

 

SYSTEM) requires continued operation

on power supply

40% U T

 

during power mains interruptions, it is

input lines

(60% dip in U T)

Not applicable

recommended that the (ME EQUIPMENT or

 

for 5 cycles

 

ME SYSTEM) be powered from an

IEC 61000-4-11

70% U T

 

uninterruptible power supply or a battery.

 

 

 

 

(30% dip in U T)

 

 

 

for 25 cycles

 

 

 

<5% U T

 

 

 

(95% dip in U T)

 

 

 

for 5 s

 

 

Power frequency

3 A/m

Not applicable

Power frequency magnetic fields should be at

(50/60 Hz)

 

 

levels characteristic of a typical location in a

magnetic field

 

 

typical commercial or hospital environment.

IEC 61000-4-8

 

 

 

NOTE U T is the a.c. mains voltage prior to application of the test level.

4

Page 4
Image 4
Drive Medical Design 2850-18, 2850-20 manual Forward