A.Pycnometer Not Getting Vacuum

International

A.Pycnometer Not Getting Vacuum

1.Check 'O' ring to make sure it is clean and in good shape. If not, contact ELE/Soiltest for a replacement 'O' ring.

2.Check to make sure the water inlet valve is on tight.

3.Tighten the stopcock nut (at bottom of stopcock) of the water inlet valve.

B.Air Bubbles Entering from the Water Inlet Valve

1.Check to make sure the water inlet valve is on tight.

2.Tighten the stopcock nut (at bottom of stopcock) of the water inlet valve.

C.Clogged In-Line Filter Replace the filter assembly.

D.Too Many Air Bubbles in Water

1.Air bubbles come from dissolved air in water. Check for this by opening water inlet valve (keeping the water inlet hose connected to the water supply) after you have gone through Step 8 of the operating procedure. The air bubbles should dissolve back in water at atmospheric pressure.

2.Try filling the pycnometer with boiled water (after it has been cooled).

If problem still persists, call ELE/Soiltest Service Department for assistance.

VII.

ACCESSORIES/OPTIONS

 

 

 

 

 

 

 

 

 

 

 

 

5

 

 

ELE International, Inc.

 

Distributor:

 

ELE International, Ltd.

 

 

 

 

 

 

Chartmoor Road Chartwell Business

 

 

 

 

PO Box 608

 

 

 

 

 

 

 

 

 

 

Park

 

 

 

 

Loveland, CO 80539

 

 

 

 

 

 

 

 

 

 

Leighton Buzzard, Bedfordshire

 

 

 

 

USA

 

 

 

 

 

 

 

 

 

 

LU7 4WG England

 

 

 

 

Phone: (970) 663-9780

 

 

 

 

 

 

 

 

 

 

Phone: + (0) 1525 249200

 

 

 

 

Fax: (970) 663-9781

 

 

 

 

 

 

 

 

 

 

Fax: + (0) 1525 249249

 

 

 

 

E-Mail: soiltest@eleusa.com

 

 

 

 

 

 

 

 

 

 

E-Mail: ele@eleint.co.uk

 

 

 

 

Website: www.eleusa.com

 

 

 

 

 

 

 

 

 

 

Website: www.eleint.co.uk

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 7
Image 7
Ele 45-9300 manual Accessories/Options, A. Pycnometer Not Getting Vacuum