VReplace fuses only with ones having the same ratings for blow characteristics,
current and voltage.
VDo not remove the cover; there are no serviceable parts inside the unit. Refer
all service to authorized personnel.
VAlways check time and pressure settings before each treatment.
CAUTIONS
Federal Law (USA) restricts this device to use by, or at the direction of a physician.
Position the CoughAssist MI-E so that the air intake ports on the side and
rear of the unit are not blocked.
Never operate the device unless a bacterial/viral filter is attached to the
patient circuit.
Always use a new filter when using the device on a new patient.
This device is designed for Intermittent Operation Only and not for continuous
use. The device should not be cycled continuously for more than 5 minutes.
After such time, the unit should either be turned off or left idling with the
blower on for at least 5 minutes.
Turn the unit off when not in use.
Keep the cord away from heated surfaces.
Do not sterilize with ethylene oxide gas or steam sterilize the pump or
pump housing.
This device should only be used by trained personnel.
5
WARNINGS AND CAUTIONS
92804 3/28/01 8:35 AM Page 7