·On: The alarm message of the signal parameter selected will only appear as long as the signal will exceed the allowed level. The alarm message will disappear, if the value is again in the range allowed.

·On (Hold Alarm): The alarm message for the signal parameter selected will appear after the signal will have exceeded the allowed value. The alarm message will also keep alive, if the value will be back again in the range allowed.

To delete the message, you have to go back to the menu "Alarms Setup" with the F4 key or the -key and to start the function "Acknowledge alarms !".

Options:

¨F3 (Channel):

To change to the setup menu of the other channels of an MLT analyzer or analyzer module.

¨F5 (View...):

To change to the submenu "Parameter (i.e.: Concentration) Alarm Status", where you will find the alarm status ("On/Off") of the four alarms for the parameter selected.

5 - 24

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98