5)Calculate linearization curve:
·Change in the menu "Linearization" to the menu line "Calculate linearization curve coefficients..." with the −
·Press the 8
TAG | 3163.860 ppm |
Used method:
All ranges
** Unused
Used |
|
|
Used method: |
| Splines |
Measure |
| Back... |
F1 | F3 | F4 |
·Press the 8
·Change to the line "Effected range for
·Press the 8
·Select any of the four ranges or all ranges with the −
·Change to the line "Start calculation for used method !" with the −
6)Enable the linearization:
·Change in the menu "Linearization" with the −
·Press the 8
·Confirm your adjustment with the 8
7)Linearization control:
·Repeat the measurement with the same setpoint values used for the linearization curve.
·The full scale deviation has to be better than 1% rel. !
Further Setups and ramifications in the menu "Linearization":
¨Set all raw/setpoint values to 0.0 !
Starting this function will reset all
¨Menu Line "Linearization Procedure...":
Via this line you can change to the menu "Linearization Procedure", where you can
·set up gas blender parameters
·set up the valve position
·control the linearization procedure status
5 - 36 | NGA 2000 |
90003482(2)