5.1 Analyzer Module Setup

5.1.6 Programmable Locic Control (PLC)

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

Programmable logic control (PLC)

TAG

95.00 ppm

-- Programmable Logic Control --

Programming...

Disabled

Measure

 

Back...

F1

F2

F4

In the menu "Programmable Logic Control" you can enable or disable the logic control of the MLT analyzer, of the analyzer module or of the digital outputs by a corresponding program. So, you can decide for instance if some calibration results should be sent to a certain analyzer output. If you want to realize a program, you have to change to the submenu "Program" (line "Programming..."). If you want to control the output status you have to change to the submenu "PLC Outputs" (line "Results...").

Note:

The PLC is only able to work with signals of the MLT analyzer or analyzer module selected! It is not possible to work with auxiliary modules or external undefined input signals!

Set up parameters:

¨Select the line of variables or any menu line with the -key or the -key.

¨Select the variable or change to the submenu with the 8 -key or the -key.

¨Select "Enabled" or "Disabled" in the line "PLC is" with the -key or the -key.

¨Confirm the parameter selected with the 8 -key or cancel and go back to the last value with the F2 key.

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

5 - 37