NGA-2000 System Calibration

3.4.3 Control System Calibration by Programmable Inputs

We can use the possibility to control SYSCAL by triggering with programmable inputs. This feature is supported by inputs of the DIO-Module.

Assigning of an programmable input can be accomplished by using the menus for DIO module.

Analyzer and I/O-module expert configuration...

I/O module controls...

DIO module(s)...

Inputs...

-- DIO- Module Inputs --

Inputnumber:

5

DIOINPNUMC

Choose module...

 

 

Choose function...

 

 

_______________________________________________

 

Slot ID:

1

DIOSLOTIDC

Signal name:

SYS:Cancel-Cal

DIOINPSIGC

Signal level:

OFF

DIOINPSTATC

Signal comes from:

Control Module

DIOINPSRCC

There we have to select

1.wanted input number

2.the appropriate module as Source Module (see following table)

3.the wanted function for system calibration

The provided functions for system calibration are:

function

source

function

function

enum-value

 

module

(positive edge)

(negative edge)

in STINAME

SYS:Zero-Cal

CM

start system zero-calibration

-

6

SYS:Zero/Span-Cal

CM

start system zero/span-calibration

-

7

SYS:Program-Cal

CM

start user defined sequence program

-

8

 

 

calibration

 

 

SYS:Cancel-Cal

CM

stop a running procedure and disable

enable starting

9

 

 

any start commands

commands

 

SYS:CAL-Test-Mode

CM

switches into test mode

switches off test mode

10

SYS:AM-Zero-Gas

AM

switches zerogas-valve

stop a running procedure

11

SYS:AM-Span-Gas1

AM

switches spangas-valve for range 1

stop a running procedure

12

SYS:AM-Span-Gas2

AM

switches spangas-valve for range 2

stop a running procedure

13

SYS:AM-Span-Gas3

AM

switches spangas-valve for range 3

stop a running procedure

14

SYS:AM-Span-Gas4

AM

switches spangas-valve for range 4

stop a running procedure

15

Please take notice that all actions are edge-triggered. Therefore take care of functionality of positive as well as negative edge.

Supplement - 18

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98