3.4.3 Control System Calibration by Programmable Inputs
We can use the possibility to control SYSCAL by triggering with programmable inputs. This feature is supported by inputs of the
Assigning of an programmable input can be accomplished by using the menus for DIO module.
Analyzer and
↓
I/O module controls...
↓
DIO module(s)...
↓
Inputs...
↓
Inputnumber: | 5 | DIOINPNUMC |
Choose module... |
|
|
Choose function... |
|
|
_______________________________________________ |
| |
Slot ID: | 1 | DIOSLOTIDC |
Signal name: | DIOINPSIGC | |
Signal level: | OFF | DIOINPSTATC |
Signal comes from: | Control Module | DIOINPSRCC |
There we have to select
1.wanted input number
2.the appropriate module as Source Module (see following table)
3.the wanted function for system calibration
The provided functions for system calibration are:
function | source | function | function | |
| module | (positive edge) | (negative edge) | in STINAME |
CM | start system | - | 6 | |
CM | start system | - | 7 | |
CM | start user defined sequence program | - | 8 | |
|
| calibration |
|
|
CM | stop a running procedure and disable | enable starting | 9 | |
|
| any start commands | commands |
|
CM | switches into test mode | switches off test mode | 10 | |
AM | switches | stop a running procedure | 11 | |
AM | switches | stop a running procedure | 12 | |
AM | switches | stop a running procedure | 13 | |
AM | switches | stop a running procedure | 14 | |
AM | switches | stop a running procedure | 15 |
Please take notice that all actions are