NGA-2000 System Calibration

4.2.1 Filling sequence buffer

The contents of the sequence buffer depends on

calibration types and belonging module type

coherence of modules and system valves

An evaluation procedure has the choice of listed action types.

Calibration types

module type

 

 

 

action types

action data

NOP

ALL-AMs

 

 

 

USER-STEP

Nr

 

ZERO-CAL

single AM

 

 

 

SWITCH_VALVE

mask

 

SPAN-CAL

 

 

Evaluation-

 

PURGEWAIT

time

 

ZERO/SPAN-CAL

 

 

procedure

 

ZERO

AM

 

SPAN1-CAL

 

 

 

 

SPAN

AM

range

SPAN2-CAL

 

 

 

 

CALWAIT

AM

 

SPAN3-CAL

 

 

 

 

 

 

 

SPAN4-CAL

 

 

 

 

 

 

 

END-OF-PGRM

 

 

 

 

 

 

 

One calibration type has following sequence frame:

1.SWITCH_VALVE

2.do different PURGEWAIT, ZERO, or SPAN which are possible with this valve adjustment (order is determined by shortest purge time)

3.do different CALWAIT (wait for finishing a started calibration)

The evaluation procedure optimizes the order of actions by time.

The only condition is in case of ZERO/SPAN-CAL, where an AM has to do first its "ZERO" before any of its "SPAN"s can be done.

The PURGEWAIT-delay-measurement is started with the last SWITCH_VALVE-action.

We can see, that the SYSCAL-types "ZERO_ALL" and "ZEROSPAN_ALL" are special cases of a user program.

ZERO_ALL:

ZERO-CAL of ALL-AMs

ZEROSPAN_ALL:

ZERO/SPAN-CAL of ALL-AMs

Supplement - 26

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98