4.2.3.3 Zero Calibration
Before starting of zero calibration there are some parameters
∙ | CONTROL: | This variable determines which instance is controlling range of an |
|
| AM. The parameter is set to "controlled by control module" |
∙ | ZERORNGS: | This variable determines if all ranges are zeroed together or |
|
| separately. The parameter is set to "zero all ranges together". |
∙ | AMSERPHYSTAT: | This variable exists only in a |
|
| determines if the AM has a serial/parallel gas flow. |
|
| The parameter is set to "parallel gas flow" because only this state |
|
| allows SYSCAL to calibrate the different channels simultaneously. |
After modifying these parameters zero calibration is started by
∙ AMFN: | This variable can start functions on a AM. It is set to the value for |
| function "Zero". |
4.2.3.4 Span Calibration
The span calibration can be done only in single ranges. Any action for a span calibration in a range can only be done if "spangas value <> 0.0".
Before starting then span calibration there are some parameters
∙ | CONTROL: | This variable determines which instance is controlling range of an |
|
| AM. The parameter is set to "controlled by control module" |
∙ | CALRANGES: | This variable determines if all ranges are spanned together or |
|
| separately. The parameter is set to "span all ranges separately". |
∙ | AMSERPHYSTAT: | This variable exists only in a |
|
| determines if the AM has a serial/parallel gas flow. |
|
| The parameter is set to "parallel gas flow" because only this state |
|
| allows SYSCAL to calibrate the different channels simultaneously. |
∙ | CRANGE: | This variable controls the current range of an AM. It is set to the |
|
| respectively required range. |
After modifying these parameters span calibration is started by
∙ AMFN: | This variable can start functions on a AM. It is set to the value for |
| function "Span". |