TAG

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

37.50

 

 

 

 

 

 

 

 

 

 

 

 

ppm CH4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.00

 

 

 

 

Range: 1

50.00

 

 

 

 

 

 

 

 

 

 

Failures:

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Maintenance-Requests: No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature:

 

20.0 C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.0

 

 

 

 

 

100.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Operation:

 

 

 

Ready

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display

 

Status...

 

Main...

 

Channel

 

Calib...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

 

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

 

37.50 ppm

 

 

 

 

 

 

 

 

 

 

 

-- Analyzer Module Calibration --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calibration procedure status...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Start zero calibration procedure !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Start span calibration procedure !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Time & Date:

 

 

 

12:50:55 June 16, 1998

 

 

 

 

 

 

 

 

Valve position:

 

 

 

 

 

 

Samplegas

 

 

 

 

 

 

 

 

Expected zero gas:

 

 

 

 

 

 

0.00 ppm

 

 

 

 

 

 

 

 

Expected span gas:

 

 

 

 

 

 

50.00 ppm

 

 

 

 

 

 

 

 

Span gas name:

 

 

 

 

 

 

 

Methane

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

Status...

 

Channel

 

Back...

 

Valves...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

 

37.50 ppm

 

 

 

 

 

 

 

 

 

 

-- Analyzer Module Calibration --

 

 

 

 

 

 

 

 

 

 

Calibration procedure status...

 

 

 

 

 

 

 

 

 

 

 

 

 

Start zero calibration procedure !

 

 

 

 

 

 

 

 

 

 

 

 

 

Start span calibration procedure !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Time & Date:

 

 

 

12:50:55 June 16, 1998

 

 

 

 

 

 

 

Valve position:

 

 

 

 

 

 

Samplegas

 

 

 

 

 

 

 

Expected zero gas:

 

 

 

 

 

 

0.00 ppm

 

 

 

 

 

 

 

Expected span gas:

 

 

 

 

 

 

50.00 ppm

 

 

 

 

 

 

 

Span gas name:

 

 

 

 

 

 

 

Methane

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

Status...

 

Channel

 

Back...

 

Valves...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.6Span Calibration

Change to the menu "Analyzer Module Calibration"

F5

Press

Caution:

Before starting span calibration, realize that spangas with correct concentration is available ! (See also section 5.1.1, p.5-5... !)

Change to the line "Start span calibration procedure !"

Press twice to get the line "Start span calibration procedure !" white on black.

Notes:

·Normally, all measurement ranges of the same chan- nel will be calibrated simul- taneously.

·To calibrate seperately you have to change the para- meters (see 5.1.1 p. 5-11).

Start the

span calibration

Press

or

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

4 - 43