3 Display and Keyboard

3.7 The STATUS Key

From the single component display or from the main menu you can change with the F2 key (STATUS) to the menu "Current measurement parameters":

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

Current measurement parameters

 

 

 

 

 

 

 

 

 

 

 

Analyzer gas measured:

 

 

 

 

 

 

 

NOx

 

 

 

 

 

 

 

 

Measurement range number:

 

 

 

3

 

 

 

 

 

 

 

 

Range change control:

 

 

 

 

 

 

LOCAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Linearization mode:

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer operational state:

 

 

 

STANDBY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer alarm state:

 

 

 

 

 

 

NORMAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Alarms reported:

 

 

 

 

 

 

FAILURE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

MORE

 

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

95.0 ppm

 

Current measurement parameters

Response time:

 

 

3.0 s

Bypass flow:

 

 

1100 ml/min

Capillary pressure:

 

 

340 hPa

Detector temperature:

 

 

0.5 C

HOME

ESCAPE

 

MORE

INFO

F1

F2

F3

F4

F5

In the menu "Current measurement parameters" you can control the status of the CLD analyzer module. With the F3 key (MORE) you can change to a further menu page of this menu. With the F2 key (ESCAPE) you can always go back to the main menu. From the second menu page you can return to the first with the F4 key (MORE) or the -key.

In the menu "Current measurement parameters" you will only find status informations. You cannot change the setups there. If you want to modify some parameters, you have to change to the menu "Basic Controls" (see chapter 4) or to the menu "Analyzer module set up" (see chapter 5.1) resp. "Expert analyzer controls" (see chapter 5.5).

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

NGA 2000

3 - 5