6.3.2.9 Diagnostic Menus

Analyzer start up

Main Menu — Technical level configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Diagnostic menus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module diagnostics

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Start up analyzer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer start up

 

 

 

 

 

 

 

 

 

 

 

 

(Press INIT to re-initialize the EEPROM data)

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer state:

 

 

 

 

 

 

 

STANDBY

 

 

 

 

 

 

 

 

Sensor temperature:

 

 

 

 

 

 

 

27.0

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector block temperature:

 

 

 

50.3

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Target temperature:

 

 

 

 

 

 

 

51.5

C

 

 

 

 

 

 

 

 

Converter temperature:

 

 

 

302.0

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Target temperature:

 

 

 

 

 

 

 

300.0

C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

 

REBOOT

 

INIT.

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

 

F3

 

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Analyzer start up" you will find the actual and target temperatures of the converter and of the detector block. Besides, you can reboot the analyzer module and reinitialize the EPROM in this menu:

1)Reboot:

·To restart the analyzer you have to press the F3 key.

·This reboot will start your analyzer in the same way as you do by switching on.

·If you flow the analyzer with zero gas during this time: The NO signal should be "0".

2)Re-initializing of the EPROM:

·If you press the F4 key you will update the EPROM data.

·Caution:

The analyzer will go back to the default values using this function ! After that you have to do a new start up, completely !

6 - 42