6.3.2.10 Diagnostic Menus

Nox Converter Efficiency

Main Menu — Technical level configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Diagnostic menus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module diagnostics

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOx converter efficiency

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

NOx converter efficiency

 

 

 

 

 

 

 

 

 

 

 

 

Converter temperature:

 

 

 

 

 

 

362.0 C

 

 

 

 

 

 

 

 

Converter set point:

 

 

 

 

 

 

360.0 C

 

 

 

 

 

 

 

 

 

Capillary pressure:

 

 

 

 

 

340.0 hPa

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measurement mode:

 

 

 

 

 

 

 

NOx

 

 

 

 

 

 

 

 

Measurement:

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

Last temperature reading:

 

 

 

 

361.7 C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Last measurement (other mode):

 

 

 

0.000 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

NO/NOx

 

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

F3

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In this menu you can optimize NOx converter efficiency.

1)Set up the desired converter temperature (set point):

·Press the ENTER key or the -key to select the parameter

·Select any digit you want to change with the -key or the -key and adjust a new value with the -key or the -key.

·Let the temperature stabilize.

2)Apply a test gas:

·Apply a test gas and select NO or NOx with the F3 key:

The component selected will appear in the line "Measurement mode".

3)Store the reading and the temperature:

·Press the F3 key (NO/NOx) again after the reading will be stabilized:

The current signal and the actual converter temperature will be stored automatically and displayed in the line "Last measurement" resp. "Last temperature reading".

4)Optimize the temperature:

·Compare the stored values with the actual values in the lines "Measurement" and "Converter temperature".

·Repeat the steps 1) to 3) to get the optimized converter temperature you want: Change the converter set point as desired, let it stabilize, apply test gas, F3, ...

Note: The steps 2) to 4) will only be available using the CLD software revision 2.3 !

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

NGA 2000