6.3.2.10 Diagnostic Menus
Nox Converter Efficiency
Main Menu — Technical level configuration
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
| Diagnostic menus | |||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| Analyzer module diagnostics | |||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
| NOx converter efficiency | |||||||||||||||||
|
|
|
|
|
|
|
|
| ↓ |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CLD |
|
|
|
|
|
|
|
|
|
| 95.0 ppm |
|
|
|
|
|
|
| |||
|
|
|
|
| NOx converter efficiency |
|
|
|
|
|
|
|
|
|
|
| ||||||
| Converter temperature: |
|
|
|
|
|
| 362.0 C |
|
|
|
|
|
|
| |||||||
| Converter set point: |
|
|
|
|
|
| 360.0 C |
|
|
|
|
|
|
|
| ||||||
| Capillary pressure: |
|
|
|
|
| 340.0 hPa |
|
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
| Measurement mode: |
|
|
|
|
|
|
| NOx |
|
|
|
|
|
|
| ||||||
| Measurement: |
|
|
|
|
| 95.00 ppm |
|
|
|
|
|
|
| ||||||||
| Last temperature reading: |
|
|
|
| 361.7 C |
|
|
|
|
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
| |||||||||||
| Last measurement (other mode): |
|
|
| 0.000 ppm |
|
|
|
|
|
|
| ||||||||||
|
|
|
|
|
|
|
|
|
|
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| HOME |
| ESCAPE |
| NO/NOx |
|
|
|
| INFO |
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||||
| F1 |
| F2 | F3 | F4 |
| F5 |
| ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In this menu you can optimize NOx converter efficiency.
1)Set up the desired converter temperature (set point):
·Press the ENTER key or the →
·Select any digit you want to change with the ←
·Let the temperature stabilize.
2)Apply a test gas:
·Apply a test gas and select NO or NOx with the F3 key:
The component selected will appear in the line "Measurement mode".
3)Store the reading and the temperature:
·Press the F3 key (NO/NOx) again after the reading will be stabilized:
The current signal and the actual converter temperature will be stored automatically and displayed in the line "Last measurement" resp. "Last temperature reading".
4)Optimize the temperature:
·Compare the stored values with the actual values in the lines "Measurement" and "Converter temperature".
·Repeat the steps 1) to 3) to get the optimized converter temperature you want: Change the converter set point as desired, let it stabilize, apply test gas, F3, ...
Note: The steps 2) to 4) will only be available using the CLD software revision 2.3 !
90003749(1)
NGA 2000 | 6 - 43 |