System Calibration

3.4.5 Time controlled System Calibration

To activate a time controlled system calibration we have to set-up this in a own menu page.

Expert controls and set up...

System Calibration...

Times for interval operation...

Times for Interval Operation

 

Zero-Cal is:

Disabled

Start of interval time:

16 h

Interval time:

800 h

Zero/Span-Cal is:

Enabled

Start of interval time:

10 h

Interval time:

1600 h

Program-Cal is:

Disabled

Start of interval time:

10 h

Interval time:

50 h

Next calibration events...

 

 

 

SCSTZERO

SCBGNZERO SCIVZERO SCSTZEROSPAN SCBGNZEROSPA SCIVZEROSPAN SCSTPRGM SCBGNPRGM SCIVPRGM

For the 3 different SYSCAL-modes there is the possibility to

enable/disable the activation

determine start of time calculations (start time of current day)

determine in which time intervals after start time the activation is done

The logic for the time calculations is as follows.

The basic is the respectively parameter „Start of interval time“ SIV( -time). By changing this SIV-time or changing the „interval time“IV( -time) the basic is new calculated as date of current day and time of the SIV-Time (only entire hours possible). This basic stays valid (also after a program reset!) until new values for SIV-time or IV-time are entered.

Basic-time = SIV-Time of current day

The next start of SYSCAL (ACT-time) happens at

ACT-time = Basic-time + IV-time

Is the calculated ACT-time before present point of date/time then there are added further IV-times until it is later.