System Calibration
3.4.5 Time controlled System Calibration
To activate a time controlled system calibration we have to
Expert controls and set up...
↓
System Calibration...
↓
Times for interval operation...
↓
Times for Interval Operation |
|
Disabled | |
Start of interval time: | 16 h |
Interval time: | 800 h |
Enabled | |
Start of interval time: | 10 h |
Interval time: | 1600 h |
Disabled | |
Start of interval time: | 10 h |
Interval time: | 50 h |
Next calibration events... |
|
|
|
SCSTZERO
SCBGNZERO SCIVZERO SCSTZEROSPAN SCBGNZEROSPA SCIVZEROSPAN SCSTPRGM SCBGNPRGM SCIVPRGM
For the 3 different
∙enable/disable the activation
∙determine start of time calculations (start time of current day)
∙determine in which time intervals after start time the activation is done
The logic for the time calculations is as follows.
The basic is the respectively parameter „Start of interval time“ SIV(
The next start of SYSCAL
Is the calculated
90003749(1) | NGA 2000 | Supplement - 21 |