System Calibration
3.5 Calibrate single Analyzers
Each specific analyzer has still the possibility to start a calibration over other instances than SYSCAL. For this functionality we now have some additional considerations.
∙Do not start a calibration of a single AM during a running SYSCAL. This will confuse the valve switching and the calibration commands of SYSCAL. Therefore look also for all automatic start instances of an analyzer.
∙A single calibration cannot use the purge times of system valve settings. Instead of we have to setup the AMs own parameters in the manner to wait for the required purge time after a valve switching.
∙It is refused to start single calibration of a second AM if the needed calibration gas valve is also any calibration gas valve of the first started AM.
These considerations are valid for all AMs which are included into SYSCAL.