Model NGA2000 NDIR
Instruction Manual
a.General
| Measurement Species:.................. | Heteroatomic gases such as ammonia (NH3), carbon dioxide |
|
| (CO2), carbon monoxide (CO), carbon monoxide + carbon dioxide |
|
| ethylene (C2H4), hexane (C6H14), methane (CH4), nitric oxide (NO) |
|
| and sulfur dioxide (SO2) |
| Ranges: ......................................... | 10 ppm fullscale to 100% fullscale |
|
| fullscale selections, including suppressed zero ranges |
| Repeatability:................................. | ±1% of fullscale (at constant temperature) |
| Minimum Detectable Level: ........... | 0.1% CO2 (at 1 atm. sample pressure; application dependent) |
| Noise: ............................................ | <1% of fullscale, |
| Linearity: ........................................ | ±1% of fullscale with 4th order polynomial |
| Response Time: ............................ | .05 to 30 seconds (selectable) for 0 to 90% of fullscale |
| Drift (Zero and Span):.................... | <±1% of fullscale/24 hours at constant temperature (application |
|
| dependent); |
|
| <±2% of fullscale/week at constant temperature (application |
|
| dependent) |
| Effect of Temperature:................... | <±1% of fullscale over any 10°C interval for rate of change no |
|
| greater than 10°C per hour (application dependent) |
| Environment: ................................. | Location - Class B controlled, indoor, |
| Ambient Temperature:................... | 0 to 45°C (32 to 113°F) |
| Effect of Flow:................................ | <1% of range when sample flow rate is changed by ≤250 ml/min. |
|
| (No effect if flow rate is between 0 and 500 ml/min.) |
| Power Requirements:.................... | 24 VDC ±5%, 100 W max.; ripple and noise: <100 mV |
|
| peak; line and load regulations: <±1% |
b. | Physical |
|
| Case Classification:....................... | General purpose for installation in |
| Dimensions:................................... | See Outline and Mounting Dimensions, Figure |
| Weight: .......................................... | Standard: 11 kg (24.2 lbs.); extended: 12.5 kg (27.5 lbs.) |
| Mounting:....................................... | Inside a Platform or |
| Maximum Length of LON Cable:... 1600 m (1 mile) between Analyzer Module and Platform | |
c. | Sample |
|
| Temperature:................................. | |
| Flow Rate: ..................................... | 500 to 1400 ml/min. |
| Pressure: ....................................... | Maximum 690 |
|
| cell applications |
| Particulates:................................... | filtered to <2 microns |
| Dewpoint:....................................... | <40°C (104°F), no entrained liquid |
| Materials in Contact with Sample:. Gold plated Pyrex, sapphire, quartz, Irtran, FEP Teflon, | |
|
| 316 stainless steel |
See the Preface section of the Platform manual for specifications regarding
Rosemount Analytical Inc. A Division of Emerson Process Management | Description and Specifications |