Model NGA2000 NDIR
Instruction Manual
CELL LENGTH |
| CELL VOLUME | TOTAL VOLUME | TIME FOR 2 VOLUMES | |
| CC | CC | @ 2 SCFH (1L/MIN) | ||
|
|
| |||
mm | inch | without inlet tube | cell with inlet tube | at 750 mm Hg | |
|
|
|
|
|
|
3 | 0.118 |
| 0.85 | 12 | 2 sec. |
|
|
|
|
|
|
4 | 0.157 |
| 1.14 | 12 | 2 sec. |
|
|
|
|
|
|
8 | 0.315 |
| 2.28 | 13 | 2 sec. |
|
|
|
|
|
|
16 | 0.630 |
| 3.56 | 16 | 2 sec. |
|
|
|
|
|
|
32 | 1.25 |
| 9.12 | 20 | 2 sec. |
|
|
|
|
|
|
64 | 2.52 |
| 18.24 | 25 | 3 sec. |
|
|
|
|
|
|
128 | 4.03 |
| 35.48 | 44 | 3 sec. |
|
|
|
|
|
|
232 | 9.13 |
| 65.12 | 73 | 6 sec. |
|
|
|
|
|
|
343 | 13.50 |
| 97.76 | 105 | 13 sec. |
|
|
|
|
|
|
381 | 15.00 |
| 108.60 | 116 | 14 sec. |
|
|
|
|
|
|
| Table | Cell Purging Times at Atmospheric Sample Pressure |
FAN
OUT | OUT |
| |
SAMPLE | REFERENCE |
IN | IN |
10 PSI MAX | 10 PSI MAX |
(69 kPa MAX) | (69 kPa MAX) |
PURGE IN
15 PSI (103 kPa) MAX
Note: Reference and purge gas connections are applicable only to certain applications.
Figure
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |