Instruction Manual

748332-F April 2003

Model NGA2000 NDIR

Controller Board and the network "self-install" (bind together) during initial startup. If the system contains more than one Analyzer Module, the startup procedure will interrogate the network to locate and identify all components on the network. The user will have to bind appropriate combinations of components after the startup sequence.

After the warm-up period (about one hour for the NDIR Analyzer Module), all modules are completely functional.

Check the tune and detector signal values against the factory settings listed in the

Diagnostic Service menus. If both settings are within ±5% tolerance of factory setting, go to Section 3-4 below for binding and Section 3-5 on page 3-4 for calibration. If not, refer first to Section 4-11 on page 4-6 for instructions about oscillator tune/source balance shutter adjustments.

3-4 BINDING

To achieve full coordination between Analyzer Modules and associated I/O Modules, the user must bind those components together in the System Set Up portion of the Technical Configuration Menu in software.

Analyzer PQ 322-14

23.2 % CO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0 ppm

 

 

 

50

 

 

 

 

 

 

 

 

Secondary Variable: XXXX

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Secondary Variable: XXX

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Secondary Variable: XXXX

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Secondary Variable: XXXX

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display

Parms.

 

Menu

 

 

Dual

Info

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

 

F2

 

 

F3

 

 

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3-1.

Run Mode Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

23.2 % CO

 

 

 

 

 

 

 

Analyzer XXXXXXXX

 

 

 

 

 

 

 

Main Menu

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Basic Controls

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Expert controls and setup ...

(Operational configuration)

Technical level configuration ...

(Diagnostic and manufacturing/service)

 

Delete alarm message!

 

 

 

 

 

 

Display

Parms.

 

 

 

 

Info

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

F3

 

F4

 

F5

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3-2. Main Menu Display

3-2

Startup and Operation

Rosemount Analytical Inc. A Division of Emerson Process Management