
Instruction Manual
Model NGA2000 NDIR
The NDIR Analyzer Module can be operated in linear and
The NDIR Analyzer Module is linearized with the following
Y = Ao + A1X + A2X2 + A3X3 + A4X4
Where:
X = the normalized
AO, A1, A2, A3, A4 = linearization coefficients
Y = the normalized linear output
Linearization coefficients can be developed and stored for each range through the Expert Controls menu. The operating range is selected by entering RANGE = 1, 2, 3 or 4 in the Range Mode section that that menu.
Coefficients for each selected range are automatically used when the module is in Linearization Mode. The user instructs the Analyzer Module as to which set of coefficients are to be used for each range. Maximum dynamic range is 3:1.
When ordered, special linearization coefficients for
The operator may want the module to output measurement in engineering units (ppm). This response is linear over the operating range. The following coefficients will make no correction to the
A0 = 0.00000
A1 = 1.00000
A2 = 0.00000
A3 = 0.00000
A4 = 0.00000
To calculate linearization coefficients other than those installed at the factory, take a minimum of 11 data points. (A more accurate curve can be obtained as the user approaches 21 data points. If urgent, a curve can be created with as few as four points, but this is only a temporary fix. A more accurate curve should be created as soon as possible.)
These data points can be obtained with an accurate gas divider or other flow mixing device. Before calculating coefficients, the data must be normalized to ranges of 0 to 1 units for both percent and concentration readings. Then, the axis must be reversed as illustrated in Figure
These data points can be entered into any program capable of computing a
After taking the data points, the operator may determine coefficients for
When entering the
Startup and Operation | Rosemount Analytical Inc. A Division of Emerson Process Management |