Model NGA2000 NDIR
Instruction Manual
From MAIN MENU
⇓
Expert controls and setup
Expert analyzer controls...
Auxiliary module controls...
System set up...
Analyzer Module set up...
Auxiliary Module set up...
Local I/O set up...
For Expert analyzer
controls, Auxiliary module controls, and System set up menus, see preceding flow chard. Local I/O set up information is included in the I/O Modules manual.
Analyzer module set up
Calibration gas list...
Calibration parameters...
Gas measurement parameters...
Analyzer parameter list...
Physical measurement parameters...
Displayed physical parameters...
Analyzer tag: | IR |
Auxiliary module set up
Select an auxiliary module for set up Local I/O set up...
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Module Tag
Calibration gas list | |
Zero gas - range 1: | 0 ppm |
Span gas - range 1: | 9.3 ppm |
Zero gas - range 2: | 0 ppm |
Span gas - range 2: | 19.3 ppm |
Zero gas - range 3: | 0 ppm |
Span gas - range 3: | 101.1 ppm |
Zero gas - range 4: | 0 ppm |
Span gas - range 4: | 247.1 ppm |
Calibration... |
|
#4
Calibration Parameters
Calibration adjustment limits: enabled
Calibration averaging time: 5 Secs
Calibration failure alarm: | Yes |
Cal failure error allowed: | 5% |
Calibration time out: | 60 Secs |
Zero ranges: | TOGETHER |
Span ranges: | SEPARATE |
Measurement parameters
Linearization parameters...
Response time/delay parameters...
Range setting...
Units...
Linearization functions...
Zero/Span Calibration
See #2 in Figure
Linearization parameters
Range 1 linearizer: | Enabled |
if enabled, uses curve no.: | 1 |
Range 2 linearizer: | Enabled |
if enabled, uses curve no.: | 2 |
Range 3 linearizer: | Enabled |
if enabled, uses curve no.: | 3 |
Range 4 linearizer: | Enabled |
if enabled, uses curve no.: | 4 |
Case temperature for coefficients: 45°C Set coefficients…
Response time/delay parameters
Range 1 t90 time: | 10 | Secs |
Range 2 t90 time: | 1 | Secs |
Range 3 t90 time: | 1 | Secs |
Range 4 t90 time: | 1 | Secs |
LON update rate: | 1 per Sec | |
Output delay time: | 0 | Secs |
Range Settings
Minimum range: | 10 ppm |
Maximum range: | 5000 ppm |
Range 1 lower limit: | 0 ppm |
Range 1 upper limit: | 10 ppm |
Range 2 lower limit: | 0 ppm |
Range 2 upper limit: | 100 ppm |
Range 3 lower limit: | 0 ppm |
Range 3 upper limit: | 250 ppm |
Range 4 lower limit: | 0 ppm |
Range 4 upper limit: | 1000 ppm |
#5 |
|
|
Linearity coefficients | ||
Curve 1 (2,3,4) |
| |
AO coefficient: | ||
A1 coefficient: | 0.983751 | |
A2 coefficient: | 0.115751 | |
A3 coefficient: | ||
A4 coefficient: | ||
Curve upper limit: | 10 ppm | |
Curve |
| 10% |
Curve |
| 5% |
Status: | Enabled | |
Polynomial set up |
| |
Range to be linearized: |
| 1 |
Current span gas: | 9.69 ppm | |
Calculated polynomial order : | 4 | |
Gas values shown as: | Percent | |
Gas concentrations... |
|
|
Analyzer function: | Ready | |
Gas concentrations | ||
Point 1 |
|
|
Gas value: |
| 0 |
Raw reading: |
| 0 ppm |
Linearized values: |
| 0 ppm |
Point 2 |
|
|
Gas value: |
| 10 |
Raw reading: | 11 ppm | |
Linearized values: | 10 ppm | |
Point to be measured: | Point 1 | |
Analyzer function: | Ready | |
Midpoint correction set up | ||
Range 1 (2,3,4) |
| |
Correction: | Disabled | |
Point being measured: |
| Pt 1 |
Point 1 gas concentration: 100 ppm | ||
Point 2 gas concentration: 120 ppm | ||
Point 3 gas concentration: 150 ppm |
Dependent on which auxiliary module is present on the network.
See I/O Modules manual.
Analyzer Parameter List
Analyzer tag: | |
First line’s parameter: | Flow |
Second line’s parameter: | Flow |
Third line’s parameter: | Flow |
Fourth line’s parameter: | Flow |
Linearization parameters... |
|
Units
Gas measurement units: | ppm |
Pressure measurement units: | psig |
Temperature measurement units: | F |
Variables are still sent as the basic SI unit.
Linearization functions
Polynomial set up...
Midpoint correction set up...
Use the polynomial set up to generate a
linearizing polynomial from up to 20 gases.
Linearity coefficients
See #5 above
NEXT
Point 1 reading: | 100 ppm |
Point 2 reading: | 120 ppm |
Point 3 reading: | 150 ppm |
Span gas value: | 100 ppm |
Analyzer function: | Ready |
Analyzer Parameter List
Primary Variable Parameters | |
Control mode: | LOCAL |
Output delay time: | 0 Secs |
Range 1 upper limit: | 10 ppm |
Range 2 upper limit: | 100 ppm |
Range 3 upper limit: | 250 ppm |
Range 4 upper limit: | 1000 ppm |
Range 1 lower limit: | 0 ppm |
Range 2 lower limit: | 0 ppm |
Range 3 lower limit: | 0 ppm |
Range 4 lower limit: | 0 ppm |
Analyzer Parameter List
Primary Variable Parameters |
| |
Range 1 t90 time: | 10 | Secs |
Range 2 t90 time: | 1 | Secs |
Range 3 t90 time: | 1 | Secs |
Range 4 t90 time: | 1 | Secs |
Linearizer on range 1: | Enabled | |
Linearizer on range 2: | Enabled | |
Linearizer on range 3: | Enabled | |
Linearizer on range 4: | Enabled |
#6
Physical Measurements
Barometric pressure: | 10.13 hPa |
Sample flow: | 500 cc/min. |
Case temperature: | 45°C |
Detector temperature: | 52°C |
Pressure limits... |
|
Temperature limits... |
|
Displayed Parameters
First line’s parameter: | Flow |
Second line’s parameter: | Flow |
Third line’s parameter: | Flow |
Fourth line’s parameter: | Flow |
May be displayed on the appropriate line of the single analyzer display screen.
Pressure Limits
Sample pressure upper limit: 15 kPa
Sample pressure lower limit: 10 kPa
Temperature Limits
Case upper limit: | 60 C |
Case lower limit: | 20 C |
Detector upper limit: | 69 C |
Detector lower limit: | 55 C |
Analyzer Parameter List
Calibration Parameters
Calibration averaging time: | 5 Secs | |
Calibration failure alarm: |
| Yes |
Cal failure error allowed: |
| 5% |
Calibration time out: |
| 60 Secs |
Ranges zeroed: | TOGETHER | |
Calibrate ranges: |
| SEPARATE |
Calibration adjustment limits: ENABLED
Analyzer Parameter List
Calibration Gases |
|
Zero gas - range 1: | 0 ppm |
Zero gas - range 2: | 0 ppm |
Zero gas - range 3: | 0 ppm |
Zero gas - range 4: | 0 ppm |
Span gas - range 1: | 9.3 ppm |
Span gas - range 2: | 19.3 ppm |
Span gas - range 3: | 101.1 ppm |
Span gas - range 4: | 247.1 ppm |
Figure
Rosemount Analytical Inc. A Division of Emerson Process Management | Maintenance and Service |