Model NGA2000 NDIR
Instruction Manual
SECTION 4
MAINTENANCE AND TROUBLESHOOTING
DANGER
ELECTRICAL SHOCK HAZARD
Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
NDIR Analyzer Module components that may require cleaning or replacement include:
•all printed circuit boards
•power fuse
•the module fan
•the chopper motor
•the source
•the detector
•flow sensor
•case temperature sensor
•Analysis cells and optical components
•RTD - detector temp. controller
•thermal fuse
The only components that may require adjustment are the oscillator tune and shutter
balance adjustments. These are
if the operator has changed sources, oscillator board or detector
if the operator disassembles the bench to any degree.
Also available for maintenance adjustment through the Front Panel display and keypad are the source current, preamplifier gain (referred to as Signal Gain on the
Refer to Figure
All PCBs, except the Oscillator Board and the LON Power Board, are secured to a side of the analyzer module that folds out while interconnect wiring is still in place. Remove the securing screws and fold out the entire panel.
To remove a particular board on the