Instruction Manual
Model NGA2000 NDIR
See Figure
Open Analyzer Module front panel
Access the oscillator tune value in the Primary Variable Parameters menu.
Adjust oscillator tune knob to its maximum setting and then turn it counterclockwise to 80 % of that maximum setting.
Case Temperature Sensor is attached to the motor source assembly. To replace this sensor, cut the
Disassemble and reassemble the Detector block according to Figure
The Analyzer Module is calibrated, tuned and balanced at the factory. If the diagnostic values for oscillator tune and detector signals are within ±5 % of the factory settings in Primary Variable Parameters and Modulation Check menus), no adjustment is necessary. If not, see Figure
Open Platform front panel, if present.
CAUTION
HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is open. Dropping front panel while hand or fingers are inside either handle can cause serious injury.
Access the detector signal value in the Modulation Check menu.
Flow zero gas (nitrogen) through the sample cell until the display reading stabilizes.
CAUTION
DELICATE COMPONENTS
The shutter is a delicate component with only a
Using an appropriate screwdriver that will fit through the Source Balance Adjustment Screw Holes (see Figure
Add 0.5 V to the value obtained in Step 7 by adjusting the sample shutter adjust screw clockwise. (If this reading exceeds 1.2 V, then a cell cleaning is necessary; see Section
Rotate the shutter adjust screw clockwise until the display reads the value obtained in Step 8.
Reassemble the Analyzer Module and, if appropriate, the Platform.
Rosemount Analytical Inc. A Division of Emerson Process Management |