
Instruction Manual
Model NGA2000 NDIR
When zero gas is introduced in the SAMPLE IN inlet and the Detector signal is greater 1.2V, the cells and windows require cleaning. To do this, follow this procedure:
See Figure
Fold out the PCB panel.
Disassemble the optical bench according to Figure
Rinse the cell with acetone. If this does not remove all foreign matter, use a soft brush. Do not use any metallic object inside the cell because it will scratch the gold plating. Loss of gold plating may require cell replacement.
After all matter has been removed, rinse the cell with distilled water and allow to air dry. Do not use towels.
Inspect the cell inside by holding it up to a bright light. If particles are seen, repeat Steps 5 and 6 as often as necessary.
After cleaning cell, examine
Remove any contaminants from windows with a
Reverse disassemble instructions to reassemble the unit.
Perform a leak check.
Replace desiccant if required. (See Section
After cleaning cells, the Analyzer Module must be optically aligned if the diagnostic values for oscillator tune and detector signals are not within ±5% of the factory settings. If they are not within that range, follow Steps 7 through 10 of Section
Rosemount Analytical Inc. A Division of Emerson Process Management |